A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring

Abstract

Background

Automated measurement of blood pressure (BP) in designated BP kiosks have in recent years been introduced in primary care. If kiosk blood pressure (BP) monitoring provides results equivalent to in-office BP or daytime ambulatory BP monitoring (ABPM), follow-up of adult patients could be managed primarily by self-checks. We therefore designed a comparative trial and evaluated the diagnostic performance of kiosk- and office-based BP (nurse- versus physician-measured) compared with daytime ABPM.

Methods

A trial of automated BP monitoring in three settings: a designated BP kiosk, by nurses and physicians in clinic, and by ABPM. The primary outcome was systolic and diastolic BP, with respective diagnostic thresholds of ≥135 mmHg and/or ≥ 85 mmHg for daytime ABPM and kiosk BP and ≥ 140 mmHg and/or ≥ 90 mmHg for office BP (nurse- and physician-measured).

Results

Compared with daytime ABPM, mean systolic kiosk BP was higher by 6.2 mmHg (95% confidence interval [CI] 3.8–8.6) and diastolic by 7.9 mmHg (95% CI 6.2–9.6; p < 0.001). Mean systolic BP taken by nurses was similar to daytime ABPM values (+ 2.0 mmHg; 95% CI − 0.2–4.2; p = 0.071), but nurse-measured diastolic values were higher, by 7.2 mmHg (95% CI 5.9–9.6; p < 0.001). Mean systolic and diastolic physician-measured BPs were higher compared with daytime ABPM (systolic, by 7.6 mmHg [95% CI 4.5–10.2] and diastolic by 5.8 mmHg [95% CI 4.1–7.6]; p < 0.001). Receiver operating characteristic curves of BP monitoring across pairs of systolic/diastolic cut-off levels among the three settings, with daytime ABPM as reference, demonstrated overall similar diagnostic performance between kiosk and nurse-measured values and over the curve performance for physician-measured BP. Accuracy with nurse-measured BP was 69.2% (95% CI 60.0–77.4%), compared with 65.8% (95% CI 56.5–74.3%) for kiosk BP.

Conclusions

In this study kiosk BP monitoring was not comparable to daytime ABPM but could be an alternative to in-office BP monitoring by trained nurses. The diagnostic performance of kiosk and nurse-measured BP monitoring was similar and better than that of physician-measured BP.

Trial registration

ClinicalTrials.gov (NCT04488289) 27/07/2020.

Peer Review reports

Background

Monitoring of hypertension is a core task in primary practice. The worldwide hypertension prevalence, defined as systolic blood pressure (BP) ≥140 mmHg or diastolic BP ≥90 mmHg or use of anti-hypertensive medication, is estimated to be about 30% in adults [1] and remains a predominant cardiovascular risk factor [2, 3]. Care of hypertensive patients in accordance with current guidelines [4] imposes challenges in treatment, monitoring, and comorbidity-management strategies [5,6,7,8,9,10]. Complementary techniques for BP monitoring, such as ambulatory BP monitoring (ABPM) and automated measurement in designated BP kiosks or at home, have gradually been introduced [11,12,13,14,15,16,17]. The rationale has been to increase patient engagement and reduce a potential white-coat effect on BP monitoring. According to previous findings, there is an expected difference in systolic BP of 5–15 mmHg between manual and automated office measurement, with automated BP being lower [18]. How mean values recorded in a BP kiosk correspond to daytime ABPM or to office-measured BP has not been sufficiently evaluated in practice. Currently, both ABPM and kiosk BP monitoring are commonly offered to patients in Swedish primary care. However, standardized home monitoring of hypertension had not been implemented in our setting by the time of the current study, despite recommendations [4, 16].

ABPM has the disadvantage of being dependent on special equipment, logistic conditions, and patient acceptability [19]. If kiosk BP monitoring provides results that are equivalent to daytime ABPM values, follow-up of patients could be managed primarily by self-checks, to support the treatment decisions by GPs and other physicians.

We therefore examined data for primary care patients treated or investigated for hypertension in parallel with three different techniques: automated BP measurement in a designated BP kiosk, automated measurement in the clinic by nurses and physicians, and ABPM for 24 hours.

The main study aim was to assess BP monitoring in a kiosk for self-determined BP measurement and compare in-office BP monitoring between nurses and physicians and with daytime ABPM as reference. A secondary aim was to evaluate the diagnostic performance of kiosk-based BP and office BP compared with ABPM.

Methods

Setting and population

Study participants, all age 20 years or older, were recruited at Föllinge Health Centre (population 1500), located in a rural area in Region Jämtland Härjedalen, northern part of Sweden. Eligible study participants were primary care patients treated for essential hypertension (International Classification of Diseases 10; I10) or examined for other conditions for which blood pressure control is important, e.g. cardiovascular disease, previous ischemic stroke or transitory ischemic attack, diabetes mellitus type 2, chronic kidney disease, obesity, hypercholesterolemia, and previously untreated high blood pressure. Exclusion criteria were need for emergency care, advanced stages of disease or dementia, inability to perform self-directed BP measurement, pregnancy, atrial fibrillation, arm circumference too small or wide for the device, lack of consent, or re-entry into the same study. Enrollment was conducted from February 1, 2021, through December 30, 2022.

Enrollment procedures

All participants underwent a 12-lead electrocardiogram, biochemical panel, physical examination, and physician-measured office BP. The time span from enrollment to ABPM and the last BP measurement in the office or at the kiosk was normally within 2 days, up to a maximum of 7 days, depending on logistics (Fig. 1).

Fig. 1
figure 1

Flowchart of study procedure, Abbreviation BP; blood pressure

From the enrollment visit through the last study-related BP measurement, study participants stayed on the same medication. The ABPM protocol was available to staff and participants only after the in-office and kiosk BP measurements were completed. Otherwise, there were no blinding procedures.

Randomization

To randomize participants to kiosk and office BP measurement before ABPM was set up, sealed envelopes were used at the enrollment visit. After participants completed ABPM, the kiosk and office BP measurements were repeated in reverse order (Fig. 1).

Measurement

BP measurements were performed with the following meters: in-office, Omron M7 Inteli IT (CE. 01997), certified for arm circumferences 22–42 cm; in kiosk, Omron i-Q142 (CE. 0197), certified for arm circumferences 22–42 cm; and for ABPM, Meditech ABPM-05 Blue BP-05 (CE 0120), cuff widths 120–250 mm for 23–33 cm arm circumference and 150–330 mm for 31–40 cm arm circumference. All study equipment was validated by the department of medical technology, Östersund Hospital, before and during the trial on a yearly check-up scheduled.

Kiosk and office BP values were registered after 5 minutes of quiet rest, with the patient sitting in a chair with a backrest, arm supports, and feet on the floor, with the BP cuff on the left upper arm at the level of the heart. Three BP values were recorded at an interval of 1–2 minutes, with further measurements only if the first two values differed by > 10 mmHg. Mean BP was taken using the last two measurements, and the total mean BP was taken as the average of four measurements, two before ABPM in the BP kiosk and two after ABPM, by the nurse in-office. BP at the enrollment visit was recorded as the mean of the last two measurements in a series of three (all physician-measured BP). Trained study nurses conducted the other office BP measures. Kiosk BP was assessed in a separate kiosk for self-directed BP measurement, adjacent to the clinic. Study participants were instructed about using the device before BP measurement in the kiosk, with only one study participant and no staff present in the kiosk. The study participants were instructed to register their BP values by hand on their study forms, and a mean value was subsequently calculated.

ABPM was registered every 30 minutes from 06 A.M. until 22 P.M. and every 60 minutes at night, for a total of 24 hours. The mean value calculated from the measures taken during 06 A.M. to 22 P.M was recorded as the daytime AMBP.

Measurement of in-office BP and ABPM was performed on the left upper arm for comparison with the corresponding measurements in the kiosk, where the device was adapted for BP measurement on the left upper arm, with a button control for each measurement on the right-hand side.

Outcome measures

The primary outcome was comparative measurements of systolic and diastolic BP among in-office (including nurse- versus physician-measured BP), kiosk, and daytime ABPM values. The secondary outcome was the diagnostic performance of in-kiosk and in-office (including nurse- versus physician-measured) BP, with daytime ABPM as reference. For daytime ABPM and kiosk BP, we defined the diagnostic cut-off as ≥135 mmHg systolic and/or ≥ 85 mmHg diastolic BP. For office BP, we defined the diagnostic cut-off as ≥140 mmHg systolic and/or ≥ 90 mmHg diastolic BP.

Statistical analysis

In exploratory analyses, we evaluated the performance of ≥140 mmHg systolic and/or ≥ 90 mmHg diastolic BP in kiosk and ≥ 160 systolic and/or ≥ 100 diastolic BP in office and in kiosk, with daytime ABPM as reference. We also evaluated the diagnostic performance of office and kiosk BP measures with the total 24-hour ABPM as reference (diagnostic threshold ≥130 systolic and/or ≥ 80 diastolic for 24-hour ABPM). In addition, we conducted this analysis separately for participants with drug treatment for hypertension. The diagnostic cut-off levels for in-office and kiosk BP and ABPM were selected according to recommended reference standards [4].

For sample size calculations, the difference among BP settings was set at ≥4 mmHg systolic BP with a standard deviation (SD) of 15, nominal power 0.8, alpha 0.05, and an assumed correlation of 0.5. The calculation resulted in a necessary minimum of 113 participants. The estimated SD for systolic BP was based on previous BP registrations in the same population. To account for potential late exclusions and drop-outs, we aimed to recruit up to 150 study participants.

Standard descriptive measures, means, and proportions were presented for baseline variables, and the paired t-test (two tailed) was used for comparisons of means. The statistical significance level was set at p < 0.05. SPSS (version 25, IBM Corp) and a diagnostic test evaluation calculator (MedCalc Software Ltd.) were used for the analyses [20].

Results

Of 124 potential study participants, the final study population comprised 117 persons who completed the trial with full data sets (Fig. 2).

Fig. 2
figure 2

Flowchart of patient inclusion

The mean participant age was 67 years, and 41% were women. Eighty percent had ongoing treatment for hypertension, and the remaining were visiting for other conditions for which BP monitoring was relevant. For current medication and other participant data, see Table 1.

Table 1 Characteristics of study participants (N = 117)

Primary outcomes

Compared with daytime ABPM, mean kiosk BP was higher by 6.2 (95% confidence interval [CI] 3.8–8.6) mmHg for systolic and 7.9 (95% CI 6.2–9.6) mmHg for diastolic (p < 0.001). The mean systolic nurse-measured BP in the office was similar to daytime ABPM values (+ 2.0 mmHg, 95% CI − 0.2–4.2; p = 0.071), but the diastolic nurse-measured BP was higher by 7.2 mmHg (95% CI 5.9–9.6; p < 0.001). The mean systolic and diastolic physician-office BP measures were higher compared with daytime ABPM values, by 7.6 (95% CI 4.5–10.2) mmHg for systolic and 5.8 (95% CI 4.1–7.6) mmHg for diastolic (p < 0.001; Table 2).

Table 2 Differences in mean systolic and diastolic BP and daytime ABPMa by type of BP monitoring (N = 117)

Compared with the mean systolic kiosk BP, the systolic nurse-office BP was lower, by − 4.1 mmHg (95% CI − 5.5 to − 2.8; p < 0.001), whereas mean diastolic measurements were similar between the kiosk and office values (Table 3). The mean difference between physician-measured and nurse-measured BP was 5.5 (95% CI 3.5–7.6) mmHg for systolic (p < 0.01) and − 1.3 (95% CI − 2.8–1.5) mmHg for diastolic (p = 0.08).

Table 3 Differences between systolic and diastolic BP monitoring performed in office or in a kiosk (N = 117)

Secondary outcomes

The diagnostic performance of nurse-measured and kiosk BP monitoring was compared, with daytime ABPM as reference. Nurse-measured BP, with a diagnostic threshold of ≥140 mmHg systolic and/or ≥ 90 mmHg diastolic, had a sensitivity of 66.7% (95% CI 53.7–78.1%), specificity of 72.2% (95% CI 58.4–83.6%), false-positive rate of 12.8%, false-negative rate of 17.9%, and accuracy of 69.2% (95% CI 60.0–77.4%). Kiosk BP monitoring, with a diagnostic threshold of systolic ≥135 mmHg and/or diastolic ≥85 mmHg, had a sensitivity of 88.9% (95% CI 78.4–95.4%), specificity of 38.9% (95% CI 25.9–53.1%), false-positive rate of 28.2%, false-negative rate of 6%, and accuracy of 65.8% (95% CI 56.5–74.3%) (Table 4).

Table 4 Diagnostic performance of nurse-measured and kiosk BP values (mm Hg) with daytime ABPM reference a (N = 117)

At a higher diagnostic threshold of systolic ≥140 mmHg and/or diastolic ≥90 mmHg for the kiosk, the sensitivity decreased to 71.1%, and the specificity increased to 66.7%. The complete characteristics of the diagnostic performances are shown in Table 4. Analysis of data for participants who were on medication for hypertension (n = 94) demonstrated only minor differences compared with the complete study cohort (Supplementary Table 1).

Extremely high BP (systolic ≥160 and/or diastolic ≥100) in office or kiosk increased the diagnostic specificity, with only a few false-positive cases for kiosk BP (2/117, 1.7%), for a specificity of 96.3 (95% CI 87.3–99.6), and no false-positive cases with nurse-measured BP. This specificity came at the expense of a lower sensitivity of 19.1 (95% CI 10.3–30.9) in the kiosk and 12.7 (95% CI 5.7–24.0) for nurse-measured values, with daytime ABPM systolic ≥135 and/or diastolic ≥85 as reference.

In the complementary analyses, we noted that the mean kiosk and office BP levels were higher before compared with after ABPM, with approximate differences of 4–6 mmHg systolic and 2–3 mmHg diastolic (p < 0.001; (Supplementary Table 2). The diagnostic performance of nurse-measured and kiosk BP values with 24-hour ABPM as reference remained at about the same level as in the primary analyses (Supplementary Table 3).

Receiver operating characteristic curves for BP monitoring in the kiosk and by the nurse or physician in the office across pairs of systolic/diastolic cut-off levels (daytime ABPM as reference) demonstrated overall similar diagnostic performance for kiosk and nurse-measured values, with the curve for physician-measured values landing under (Fig. 3).

Fig. 3
figure 3

Receiver operating characteristic curves of BP monitoring, in kiosk, nurse-measured, and physician-measured across pairs of systolic/diastolic cut-off levels (daytime ABPM reference)

Discussion

This study compared BP monitoring by automated devices in a kiosk for self-determined BP measurement and in the office by nurses and physicians, with daytime ABPM as reference. We also evaluated the diagnostic performance of all three monitoring techniques separately.

Compared with ABPM, mean kiosk systolic BP was about 6 mmHg higher and diastolic BP was 8 mmHg higher. Nurse-measured systolic BP was similar to ABPM, but nurse-measured diastolic BP was higher by 7 mmHg, as was mean physician-measured BP, by about 6–8 mmHg for both systolic and diastolic values. The differences between office and kiosk BP were small, and nurse-measured systolic BP was at most 4 mmHg lower than the mean kiosk BP.

Evaluation of the diagnostic performance of kiosk and office BP revealed only a moderately high accordance with ABPM as reference. Increasing the diagnostic threshold of hypertension from 135/85 to 140/90 in the kiosk decreased the number of false positives, but at the expense of an increased number of false negatives. The moderate diagnostic accuracy of kiosk BP monitoring in our study (65.8%) was below the 77.1% reported in a recent study from 12 Kaiser Permanente primary care centers in Washington state [16]. The diagnostic accuracy of nurse-measured BP monitoring in our setting (69.2, 95% CI 60–77.4%) was similar to that of kiosk BP monitoring (65.8, 95% CI 56.5–74.3%), in contrast to the Kaiser Permanente study, with a reported accuracy of 44.4% for clinic values [16]. The receiver operating characteristic curves based on paired diagnostic thresholds of systolic and diastolic BP indicated a similar diagnostic performance between kiosk and nurse-measured BP and over the curve for physician-measured BP. In the Kaiser Permanente study, the performance of kiosk and home BP monitoring was better than office BP monitoring across a range of systolic and diastolic BP values. In our study at-home BP monitoring was not included. The reason for not including home BP was lack of standardization and experience in our population during the study period.

In the complementary analyses of participants receiving anti-hypertensive treatment (n = 94), and in the analysis using 24-hour ABPM as reference, no clear influence on overall diagnostic accuracy was noted, compared with the primary analyses. Extremely high BP readings (≥160/100) in office or in kiosk increased specificity at the expense of lower sensitivity, confirming that white-coat hypertension could be ruled out with some assurance at that level of high BP [16].

Concerning the observed difference in mean systolic and diastolic BP values before and after ABPM, this observation is in contrast to findings by Andreades et al. [21], who could not confirm an order-related effect on automated office BP readings. To avoid order-related bias from nurse-measured and kiosk BP values, in the current study, the order of types of BP monitoring was randomized and repeated in reverse after ABPM. The physician-measured BP was registered only at the enrollment visit. Conclusions related to physician-measured BP compared with the other types of BP monitoring thus should include consideration of this difference.

Circadian variations in BP seems less likely to have influenced our results. Nurse-measured BP and kiosk BP were scheduled during the same visit before set-up of ABPM and repeated in reverse order immediately after having completed ABPM. The enrollment visits were during normal office ours 08.00–17.00 without certain preference for time of the day.

Limitations

This was a single-center study with a limited sample size, which could have influenced the external validity of our findings. The BP kiosk is not a standardized concept, despite efforts to amend this issue in the last decade. A BP kiosk could be designed and equipped in various ways, all with the common purpose of offering easy access to BP monitoring [22,23,24,25]. About four out of every five study participants received treatment for hypertension, and differences between this population and the treatment-naïve participants could not be excluded. The diagnostic performance of all types of BP monitoring depends on the cut-off levels used to define hypertension, an arbitrary parameter that can vary among studies.

Firm conclusions on the external validity of findings in a single-center study should be made with caution. Compared with the population of the QregPV register of region Västra Götaland, Sweden, which allowed recruitment of 259,753 primary care patients with hypertension [26], we recruited a smaller proportion of women (41% vs 56.5%), but mean ages were similar (67.4 vs 68.7 years). In addition, less participants in our study had ongoing drug treatment for hypertension at enrollment (80.3% vs 93.8%).

The advantages of our study approach are that the study equipment was uniformly validated and handled by a limited number of experienced staff, study participants were primary care patients receiving normal care and previously registered at the clinic, and the time span from enrollment to study completion was within 1 week, with no alterations in medication during that period. The BP kiosk was in a quiet room, adjacent to the office. The necessity of standardizing kiosk BP monitoring has been emphasized previously [27].

We believe that our findings can be of use in designing the most appropriate ways to monitor BP and care for the vast number of patients with hypertension [28].

Conclusions

In this study, kiosk BP monitoring was not comparable to daytime ABPM but could be an alternative to in-office BP monitoring by trained nurses. The diagnostic performance of kiosk and nurse-measured BP was similar and better than physician-measured BP, with daytime ABPM as reference.

Availability of data and materials

The datasets obtained during the study will be available from the corresponding author on reasonable request.

Abbreviations

ABPM:

Ambulatory blood pressure monitoring

BP:

Blood pressure

CI:

Confidence interval

SD:

Standard deviation

References

  1. Mills KT, Stefanescu A, He J. The global epidemiology of hypertension. Nat Rev Nephrol. 2020;16:223–37.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Forouzanfar MH, Liu P, Roth GA, Ng M, Biryukov S, Marczak L, et al. Global burden of hypertension and systolic blood pressure of at least 110 to 115 mm hg, 1990-2015. JAMA. 2017;317:165–82.

    Article  PubMed  Google Scholar 

  3. NCD Risk Factor Collaboration (NCD-RisC). Worldwide trends in blood pressure from 1975 to 2015: a pooled analysis of 1479 population-based measurement studies with 19·1 million participants. Lancet. 2017;389:37–55. Erratum in: Lancet. 2020;396:886

    Article  Google Scholar 

  4. Williams B, Mancia G, Spiering W, Agabiti Rosei E, Azizi M, Burnier M, et al. ESC scientific document group. 2018 ESC/ESH guidelines for the management of arterial hypertension. Eur Heart J. 2018;39:3021–104.

    Article  PubMed  Google Scholar 

  5. Banegas JR, López-García E, Dallongeville J, Guallar E, Halcox JP, Borghi C, et al. Achievement of treatment goals for primary prevention of cardiovascular disease in clinical practice across Europe: the EURIKA study. Eur Heart J. 2011;32:2143–52.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Egan BM, Kjeldsen SE, Grassi G, Esler M, Mancia G. The global burden of hypertension exceeds 1.4 billion people: should a systolic blood pressure target below 130 become the universal standard? J Hypertens. 2019;37:1148–53.

    Article  CAS  PubMed  Google Scholar 

  7. SPRINT Research Group, Wright JT Jr, Williamson JD, Whelton PK, Snyder JK, Sink KM, Rocco MV, et al. A Randomized Trial of Intensive versus Standard Blood-Pressure Control. N Engl J Med. 2015;373:2103–16. Erratum in: N Engl J Med. 2017;377:2506

    Article  Google Scholar 

  8. SPS3 Study Group, Benavente OR, Coffey CS, Conwit R, Hart RG, McClure LA, Pearce LA, et al. Blood-pressure targets in patients with recent lacunar stroke: the SPS3 randomised trial. Lancet. 2013;382:507–15. Erratum in: Lancet. 2013;382:506

    Article  Google Scholar 

  9. ACCORD Study Group, Cushman WC, Evans GW, Byington RP, Goff DC Jr, Grimm RH Jr, Cutler JA, et al. Effects of intensive blood-pressure control in type 2 diabetes mellitus. N Engl J Med. 2010;362:1575–85.

    Article  Google Scholar 

  10. Hansson L, Zanchetti A, Carruthers SG, Dahlöf B, Elmfeldt D, Julius S, et al. Effects of intensive blood-pressure lowering and low-dose aspirin in patients with hypertension: principal results of the hypertension optimal treatment (HOT) randomised trial. HOT Study Group Lancet. 1998;351:1755–62.

    CAS  Google Scholar 

  11. O'Brien E, Parati G, Stergiou G, Asmar R, Beilin L, Bilo G, et al. European Society of Hypertension Working Group on blood pressure monitoring. European Society of Hypertension position paper on ambulatory blood pressure monitoring. J Hypertens. 2013;31:1731–68. Erratum in: J Hypertens 2013:2467

    Article  CAS  PubMed  Google Scholar 

  12. Parati G, Stergiou G, O'Brien E, Asmar R, Beilin L, Bilo G, et al. European Society of Hypertension Working Group on blood pressure monitoring and cardiovascular variability. European Society of Hypertension practice guidelines for ambulatory blood pressure monitoring. J Hypertens. 2014;32:1359–66.

    Article  CAS  PubMed  Google Scholar 

  13. Seidlerová J, Gelžinský J, Mateřánková M, Ceral J, König P, Filipovský J. In the aftermath of SPRINT: further comparison of unattended automated office blood pressure measurement and 24-hour blood pressure monitoring. Blood Press. 2018;27:256–61.

    Article  PubMed  Google Scholar 

  14. Seidlerová J, Ceral J, Mateřánková M, König P, Řiháček I, Vysočanová P, et al. Long-term relationship between unattended automated blood pressure and auscultatory BP measurements in hypertensive patients. Blood Press. 2019;28:34–9.

    Article  PubMed  Google Scholar 

  15. Filipovský J, Seidlerová J, Kratochvíl Z, Karnosová P, Hronová M, Mayer O Jr. Automated compared to manual office blood pressure and to home blood pressure in hypertensive patients. Blood Press. 2016;25:228–34.

    Article  PubMed  Google Scholar 

  16. Green BB, Anderson ML, Campbell J, Cook AJ, Ehrlich K, Evers S, et al. Blood pressure checks and diagnosing hypertension (BP-CHECK): design and methods of a randomized controlled diagnostic study comparing clinic, home, kiosk, and 24-hour ambulatory BP monitoring. Contemp Clin Trials. 2019;79:1–13.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Green BB, Anderson ML, Cook AJ, Ehrlich K, Hall YN, Hsu C, et al. Clinic, home, and kiosk blood pressure measurements for diagnosing hypertension: a randomized diagnostic study. J Gen Intern Med. 2022;37:2948–56.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Kjeldsen SE, Lund-Johansen P, Nilsson PM, Mancia G. Unattended blood pressure measurements in the systolic blood pressure intervention trial: implications for entry and achieved blood pressure values compared with other trials. Hypertension. 2016;67:808–12.

    Article  CAS  PubMed  Google Scholar 

  19. Thompson MJ, Anderson ML, Cook AJ, Ehrlich K, Hall YN, Hsu C, et al. Acceptability and adherence to home, kiosk, and clinic blood pressure measurement compared to 24-H ambulatory monitoring. J Gen Intern Med. 2023:1–8. https://doi.org/10.1007/s11606-023-08036-3.

  20. https://www.medcalc.org/calc/diagnostic_test.php (Version 20.218); Accessed 22 Feb 2023.

  21. Andreadis EA, Geladari CV, Angelopoulos ET, Savva FS, Georgantoni AI, Papademetriou V. Attended and unattended automated office blood pressure measurements have better agreement with ambulatory monitoring than conventional office readings. J Am Heart Assoc. 2018;7:e008994.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Alpert BS. Are kiosk blood pressure readings trustworthy? Blood Press Monit. 2012;17:257–8.

    Article  PubMed  Google Scholar 

  23. Al Hamarneh YN, Houle SK, Chatterley P, Tsuyuki RT. The validity of blood pressure kiosk validation studies: a systematic review. Blood Press Monit. 2013;18:167–72.

    Article  PubMed  Google Scholar 

  24. Alpert BS, Dart RA, Sica DA. Public-use blood pressure measurement: the kiosk quandary. J Am Soc Hypertens. 2014;8:739–42.

    Article  PubMed  Google Scholar 

  25. Padwal RS, Townsend RR, Trudeau L, Hamilton PG, Gelfer M. Comparison of an in-pharmacy automated blood pressure kiosk to daytime ambulatory blood pressure in hypertensive subjects. J Am Soc Hypertens. 2015;9:123–9.

    Article  PubMed  Google Scholar 

  26. Bager JE, Mourtzinis G, Andersson T, Nåtman J, Rosengren A, Björck S, et al. Trends in blood pressure, blood lipids, and smoking from 259 753 patients with hypertension in a Swedish primary care register: results from QregPV. Eur J Prev Cardiol. 2022;29:158–66.

    Article  PubMed  Google Scholar 

  27. Cohen DL, Townsend RR. Blood pressure readings using public kiosks or smart phone apps: caveat emptor (for now). J Clin Hypertens (Greenwich). 2017;19:946–7.

    Article  PubMed  Google Scholar 

  28. GBD 2017 Risk Factor Collaborators. Global, regional, and national comparative risk assessment of 84 behavioural, environmental and occupational, and metabolic risks or clusters of risks for 195 countries and territories, 1990–2017: a systematic analysis for the Global Burden of Disease Study 2017. Lancet. 2018;392:1923–94. Erratum in: Lancet. 2019;393:132. Erratum in: Lancet. 2019;393:e44

    Article  Google Scholar 

Download references

Acknowledgments

The authors sincerely thank the staff at the Föllinge Health Centre for their contribution.

Funding

Open access funding provided by Umea University. The study was funded by grants from the Region Jämtland Härjedalen. The funders had no role in the design of the study or collection, analysis, and interpretation of data, or in writing the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

Both authors have made substantial contributions. Conception and design: GN, AL. Acquisition of data: GN. Statistical analysis and interpretation of data: AL, GN. GN wrote the first draft, which was revised critically for important intellectual content by both authors. The authors have read and given final approval of the version to be published.

Corresponding author

Correspondence to Gunnar Nilsson.

Ethics declarations

Ethics approval and consent to participate

Ethical approval was obtained from the Swedish Ethical Review Authority (2020–01375), and the study was registered at ClinicalTrials.gov (NCT04488289) 27/07/2020. All participants provided written informed consent before enrollment.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nilsson, G., Lindam, A. A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring. BMC Cardiovasc Disord 24, 27 (2024). https://doi.org/10.1186/s12872-023-03701-1