Technical advance in silico and in vitro development of a new bipolar radiofrequency ablation device for renal denervation
BMC Cardiovascular Disorders volume 21, Article number: 500 (2021)
Abstract
Background
Renal denervation with radiofrequency ablation has become an accepted treatment for drug-resistant hypertension. However, there is a continuing need to develop new catheters for high-accuracy, targeted ablation. We therefore developed a radiofrequency bipolar electrode for controlled, targeted ablation through Joule heating induction between 60 and 100 °C. The bipolar design can easily be assembled into a basket catheter for deployment inside the renal artery.
Methods
Finite element modeling was used to determine the optimum catheter design to deliver a minimum ablation zone of 4 mm (W) × 10 mm (L) × 4 mm (H) within 60 s with a 500 kHz, 60 Vp-p signal, and 3 W maximum. The in silico model was validated with in vitro experiments using a thermochromic phantom tissue prepared with polyacrylamide gel and a thermochromic ink additive that permanently changes from pink to magenta when heated over 60 °C.
Results
The in vitro ablation zone closely matched the size and shape of the simulated area. The new electrode design directs the current density towards the artery walls and tissue, reducing unwanted blood temperature increases by focusing energy on the ablation zone. In contrast, the basket catheter design does not block renal flow during renal denervation.
Conclusions
This computational model of radiofrequency ablation can be used to estimate renal artery ablation zones for highly targeted renal denervation in patients with resistant hypertension. Furthermore, this innovative catheter has short ablation times and is one of the lowest power requirements of existing designs to perform the ablation.
Background
Hypertension increases the risk of heart disease, stroke, and death. A 12–13 mmHg reduction in systolic blood pressure reduces the risk of cardiovascular disease by at least 21% [1]. In 2020, the American Heart Association (AHA) reported that 31.9% of Americans have hypertension, resulting in a healthcare expenditure of $55.9 billion [2]. AHA/American College of Cardiology guidelines define resistant hypertension (RH) as hypertension resistant to three or more medications, resulting in higher blood pressure than a ≥ 130/80 mmHg target or a blood pressure < 130/80 mmHg in a patient taking ≥ 4 antihypertensive drugs. RH affects 12–15% of patients treated for hypertension and remains a clinical management challenge [3].
Hypertension is, at least in part, caused by chronic activation of the sympathetic nervous system. As early as 1889, experiments with dogs showed that renal or splanchnic nerves stimulation caused changes in blood pressure [4]. In 1925, Adson performed surgical sympathectomy (disconnection of the sympathetic nerves at the trunk) to treat hypertension [5]. For the subsequent two decades, surgical sympathectomy (thoracolumbar splanchnicectomy) became the procedure of choice for patients with HT who did not respond to diet or the then limited available pharmacological therapy [6]. Indeed, between 1938 and 1947, approximately 2400 patients with hypertension were treated by surgical sympathectomy [7]. Although blood pressure reductions after the operation were often significant, patients were seriously affected by intolerable post-operative postural hypotension [8]. When the first effective antihypertensive drugs (diuretics) became available in the mid-1950s, pharmacological therapy became the standard of care for patients with hypertension [9].
Since stimulation of the renal nerves raises blood pressure through vasoconstriction and volume and sodium retention [10,11,12], ablating renal nerve activity (renal denervation; RD) effectively lowers blood pressure [13]. The highest density of renal sympathetic nerves is within 12–15 mm of the abdominal aorta and within 4 mm radially from the renal artery inner lumen [14]. RD's target is to disconnect the efferent and afferent renal sympathetic nerves between the kidneys and the central nervous system to reduce blood pressure. Indeed, the first bilateral RDs for hypertension were performed in 1934 and 1935, establishing the procedure's safety with no adverse effect on renal function [15, 16].
Multiple RD methods are incorporated in devices using radiofrequency, chemicals, or ultrasound [13]. However, an RD human trial in the USA, SYMPLICITY HTN-3, did not meet the primary efficacy endpoint [17]. Still, subsequent trials using next-generation multielectrode radiofrequency (RF) and ultrasound renal nerve ablation systems have shown consistent blood pressure-reducing effects [11, 12]. The SYMPLICITY HTN-3 trial resulted in the redesign of the ablation catheter from a single point unipolar ablation electrode to a four-quadrant unipolar device.
To date, no RD device has been cleared by the Food and Drug Administration (FDA) in the USA except for investigational purposes [18]. Here we employed a computerized (in silico) finite element model (FEM) to optimize an innovative bipolar RF ablation device. The model was validated with thermochromic phantom (TCP) tissue by comparing ablation zone geometries. Our new catheter design does not block renal blood flow, does not require cooling, and has low power consumption due to the bipolar electrode design.
Methods
Computational modeling
The finite-element method (FEM) has previously been used to model RF tumor or cardiac ablation [19,20,21]. Here we also used FEM to solve electromagnetic equations coupled with bioheat equations for bipolar electrodes in contact with tissue design to deliver a minimum ablation zone of 4 mm (W) × 10 mm (L) × 4 mm (H) within 60 s with a 500 kHz, 60 Vp-p signal and 3 W maximum.
COMSOL version 5.6 (COMSOL, Stockholm, Sweden) was used to generate geometric models, assign material properties and boundary conditions, define infinite element domains, generate meshes, and perform the coupled electromagnetic heating calculations. A coupled analysis was required since the electrical properties of tissue change with temperature and, therefore, the electric field profile must be recalculated at each time step. All analyses were performed on a PC equipped with an Intel® Core™ i7-9700 CPU @ 3 GHz, 64 GB of memory, and Windows 10 Home OS.
Model geometry
Multiple software packages are commercially available to perform FEM for solving partial differential equations. We used the readily available COMSOL software to develop a 3D FEM model geometry of bipolar cylindrical electrodes inside a renal artery surrounded by connective tissue (Fig. 1A). The renal artery internal diameter was 4.1 mm [14], the wall (tunica intima and media) thickness was 0.5 mm [22], and the surrounding connective tissue was modeled to a radius of 10 mm. An infinite element domain was defined around the geometric model for approximating an infinitely large domain (Fig. 1A). The electrode dimensions were 0.61 mm outside diameter, 0.1 mm thickness, and 3 mm long assembled over a Pellethane® tube. The electrodes were designed to be in contact with the renal artery's inner wall through an exposed window of 1 mm × 2.5 mm, as shown in Fig. 1B (patent pending).
Electrical and thermal properties
Table 1 shows the material properties at 500 kHz used in the computational model to describe the blood, artery wall, connective tissue, thermoplastic catheter material (Pebax®), and electrode components.
Temperature changes result in electrical conductivity changes in biological tissues (artery walls and connective tissue) [23] and, during RF ablation, the temperature of tissues increases. The changes to the electrical conductivity in tissues included in the model were described using:
where \(\mathop \sigma \limits_{^\circ }\) is the initial electrical conductivity at the reference temperature, \(k_{1}\) is the temperature coefficient, and \(\Delta T\)[°C] is the temperature difference from the initial reference temperature. We set \(k_{1} = 2.05\%\)/°C, and the linear electrical conductivity in the model was determined by
For the artery wall and connective tissue, the equation was
Both equations are bounded within the experimental data of 30–80 °C [24].
Multi-physics computation
Electric field
When electric fields are applied to tissue, the temperature in the zone between the bipolar electrodes increases due to Joule heating through tissue conduction and dielectric loss. For the tissues in the model at 500 kHz, dielectric losses are negligible, so only the resistive loss (i.e., Joule heating) was considered in this simulation [25]. The electric field E (V/m) profile was calculated from the negative gradient of the voltage using the equation
where ∇ is the del operator that turns the scalar voltage V (V) to a vector field. Then, current density J (A/m2) was computed from
where \(\rho e\) is the electrical resistivity of the material (Ω m). Then, the average power dissipated in the tissue between the electrodes was determined by
where \(\Re\) is the real part from the vector dot product of conjugate \({\varvec{J}}\) and \({\varvec{E}}\).
Bioheat equation
Penne’s bioheat equation describes the temperature distribution increase surrounding the bipolar electrodes during the Joule heating process and incorporates the convective heat transfer from blood perfusion within the tissue
where \(\rho\) is the tissue density (kg/m3), Cp is the tissue-specific heat (J/kg K), and k is the tissue thermal conductivity (W/m K) of the components in the model. Tb is the blood temperature (assumed to be 37 °C), \(\rho_{{\text{b}}}\) is blood density (kg/m3), Cp,b the specific heat of the blood (J/kg K), and ωb is the blood perfusion (1/s). Qmet is the energy generated by metabolic processes (W/m3) and is negligible for the renal artery model. The blood temperature in large surrounding vessels has minimal impact on the thermal field between the bipolar electrodes due to their distance from the renal artery and was therefore not included in the model.
Boundary conditions
Artery walls and surrounding tissue with Dirichlet boundary conditions: temp = normal body temperature (37 °C) and voltage = 0 V.
Ablated tissue
To determine the ablation zone (i.e., where cell death occurs), the Thermal Damage transformation model was included in the COMSOL model with a temperature threshold. The isothermal surfaces were plotted using the following parameters to determine the ablation zone: damage temperature 60 °C, damage time 1 s, necrosis temperature 100 °C, and enthalpy change 0 J/kg.
In vitro experiments
We prepared a thermochromic tissue phantom (TCP) for direct visualization of ablation zone geometry and compared it with the FEM-simulated ablated tissue 3D geometries (Fig. 4A). The TCP tissue approximates the electrical properties of real tissue and provides a good in vitro setup for testing the bipolar electrodes. The TCP tissue was prepared using polyacrylamide gel with a thermochromic ink additive that permanently changes color from white to magenta when heated over 60 °C. The TCP was formulated as described by Mikhail et al. [26] and contained: deionized water 76.1 (v/v), 40% acrylamide/bis-acrylamide 17.5 (v/v), magenta MB60°C concentrate 5.0 (v/v) sodium chloride (NaCl) 0.9 (w/v), ammonium persulfate (APS) 0.14 (w/v), and N, N, N′, N′-tetramethylethylenediamine (TEMED) 0.14 (v/v). The electrical conductivity was adjusted to match the connective tissue since this is the most relevant parameter for the thermal Joule heating effect.
The TCP was ablated with a bipolar electrode modeled in Fig. 1B and connected to a waveform generator and linear amplifier, as shown in Fig. 2. The bipolar electrode inside an artery as per the FEM in silico model was approximated with the TCP tissue resting flat and the bipolar electrode on top, as shown in Fig. 2. This experimental strategy was chosen to provide a common experiment framework for application to future ex vivo modeling on a dissected and split renal artery with the bipolar electrode placed on top. The exposed surface of the electrodes was in contact with the TCP and impregnated with saline solution, which mimics the conductivity of blood. An ablation signal of 3.0 Vp-p and 500 kHz was generated with an Agilent 33220A waveform generator (Agilent, Santa Clara, CA, USA) and amplified to 60 Vp-p with a Pendulum F20A fixed-gain linear amplifier (Pendulum, Stockholm, Sweden). A Tektronix TDS2001C oscilloscope (Tektronix, Beaverton, OR, USA) and a Fluke 8846A Precision Multimeter (Fortive, Everett, WA, USA) were connected to monitor and confirm that the appropriate ablation signal was applied to the TCP through the electrodes (not shown in Fig. 2 for simplification).
Ten in vitro experiments were conducted with the same setup as in Fig. 2. The 3D effects of the RF ablation on the TCP tissue were assessed by measuring the length and width with a ruler, with subsequent transection of the specimen to measure the height. The ablated area of TCP tissue was also quantified using ImageJ [29] to analyze the ten images of the TCP ablations scaled to 86 pixels = 1 mm using the Threshold Color tool with the default thresholding method (red color) and the following settings: hue = 217, saturation = 121, brightness = 68, color space HSB, and no dark background.
Results
RF ablation signal power
Table 2 summarizes the peak power delivered during the 60-s RF ablation signal application. The correlation between the FEM in silico model and in vitro TCP tissue ablation peak power was within 5%.
RF ablation catheter design
The bipolar electrode design in this study (Fig. 1) can be assembled in a 6 Fr catheter with four splines carrying either 2 or 4 electrodes per spline. Each spline is made of thermoplastic (Pebax®) extrusions with their inner lumen containing a pre-shaped nickel-titanium (Nitinol) and electrical wires (Fig. 3). The Nitinol provides a collapsible basket to be delivered through femoral access to the renal artery. It expands to the pre-shaped form creating firm connectivity between the electrodes and the artery inner walls. Each bipolar electrode can be independently selected for ablation of each quadrant in contact with the splines inside the artery (patent pending).
In silico and in vitro comparisons
The isothermal plot from the computer model (Fig. 4A) predicted the tissue volume that reached an ablation temperature of ≥ 60 °C after 60 s of ablation due to the induced Joule heating. The ablated area consisted of two elongated lobes with an approximate maximum dimension of 4 mm × 10 mm × 4 mm (length × width × height).
The ablated TCP tissue (Fig. 4B) was measured and compared with the isothermal plot from the computer model (Fig. 4A) and is shown in Fig. 4C. Note the excellent geometrical correlation between simulated and experimentally ablated zones. For more accurate quantification, the images of ablated area of TCP tissue were quantified using ImageJ (Fig. 4D). The experimentally ablated zone measured 3.2 mm × 9.6 mm × 3.4 mm with a standard deviation of 0.12 mm for the ten experimental replicates. These values closely matched the model prediction of 4 mm × 10 mm × 4 mm, indicating that the computer model accurately predicted the RF ablation zone in vitro.
Renal artery blood flow effects
The RF ablation zones in Fig. 4 considered renal blood flow thermal dynamics. The blood temperature inside the artery was maintained at 37 °C, as shown in Fig. 5. Thermal dispersion from the positive electrode over the inside artery wall was noted in the direction of renal blood flow set at 0.7 m/s, a density of 1050 kg/m3, and normal blood pressure of 120/80 mmHg [27].
Discussion
Resistant hypertension remains common and a clinical management challenge [1]. After initially disappointing clinical trial results, the technique's refinement has now resulted in the inconsistent demonstration of RD efficacy for patients with treatment-resistant hypertension [11, 12]. Additionally, the Food and Drug Administration (FDA) has not approved any device for RD except for investigational purposes, despite the many ongoing clinical trials for RD and suggesting a need for further catheter refinement.
Several RF ablation catheter devices with the CE mark are authorized for use in the European Union (EU) (Table 3). The configuration column states the operating mode between unipolar (requiring a ground or indifferent pad) and bipolar (no need for another external pad). All devices listed are unipolar devices except for the Vessix™ (Boston Scientific, Marlborough, MA) and our device, the only bipolar device. The downside of VessixTM is the inclusion of a balloon with bipolar electrodes [18]. This has the consequent disadvantage that once inflated, the balloon blocks renal blood flow during ablation. Our electrode design based on a basket catheter does not require a balloon, so it has the advantage of no renal blockage during ablation. Furthermore, there is no need for cooling, and the catheter has lower ablation times and power requirements than most existing designs.
RF ablation devices have evolved from single unipolar electrodes (the original Symplicity HT1™, Medtronic, Santa Clara, CA, USA) to multi-electrode unipolar systems (Spyral™) [28] Here we designed a bipolar electrode for RF ablation with the electrode design, dimensions, geometry, and ablation area simulated and optimized via FEM in silico. Several software packages are available at varying costs based on the complexity and variables in the model. However, their benefit includes reducing further costs for design iterations that are simple to change in the FEM and predicting the results before designing and building the device and implementing in vitro experiments. The FEM simulation included the surrounding artery and connective tissue's electrical properties to predict the ablation zone dimensions to achieve renal nerve ablation. A TCP tissue was used to compare the in vitro ablation zone with FEM in silico simulations. A limitation of the TCP is its formulation with only one set of electrical characteristics; consequently, the experimental strategy was to define the electrical properties of the renal artery walls in direct contact with the surface of the electrodes to provide conservative experimental data on the voltage/ablation time required for further optimization when moving to ex vivo experimentation. Another limitation of the TCP is measurements of ablation zones. TCP starts changing color at 60 °C, and there is considerable visual variability to determine the exact location where the TCP starts to display temperature color transition. Nevertheless, the TCP allows visual and quantitative comparison of the geometry, shape, and form of ablation, which very closely matched the computational simulation. Therefore, this model will be helpful for further optimizing catheter design for accurate renal nerve ablation in animal models in vivo to pave the way for clinical validation.
This computational model of RF ablation can be used to estimate ablation zones in the renal artery for RD in patients with hypertension. The model provides an efficient iterative platform to design and refine RF ablation electrodes for targeted ablation geometries and dimensions. The model can be extended to other anatomical landmarks to predict the RF ablation pattern in the targeted zone. This model is complemented by the TCP to provide a relatively simple and cost-effective in vitro medium for quick feedback to the computational model from RF ablation experiments, allowing short iteration times during catheter development compared to traditional, costly, and time-consuming animal and human trials.
Conclusions
Our bipolar RF ablation electrode deserves further study and development as a potential RD device. The bipolar configuration and innovative electrode design ensure that the current density is focused on the target tissue, reducing the energy transferred to renal blood, reducing the need for cooling. Furthermore, the electrode window can be modified in shape and dimensions based on the intended location and geometry desired for ablation. This model and TCP can be used to investigate other ablation applications, e.g., tumor ablation, to design an electrode for the specifically intended zone. Further ex vivo and in vivo studies are warranted to further validate this new bipolar catheter system.
Availability of data and materials
All data generated or analyzed during this study are included in this published article.
Abbreviations
- AHA:
-
American Heart Association
- FDA:
-
Food and drug administration
- FEM:
-
Finite element modeling
- RD:
-
Renal denervation
- RF:
-
Radiofrequency
- RH:
-
Resistant hypertension
- TCP:
-
Thermochromic tissue phantom
- HT:
-
Hypertension
References
State heart disease and stroke prevention program addresses high blood pressure [https://www.cdc.gov/dhdsp/data_statistics/fact_sheets/fs_state_hbp.htm]
Virani SS, Alonso A, Benjamin EJ, Bittencourt MS, Callaway CW, Carson AP, Chamberlain AM, Chang AR, Cheng S, Delling FN, et al. Heart disease and stroke statistics-2020 update: a report from the American Heart Association. Circulation. 2020;141(9):e139–596.
Vongpatanasin W. Resistant hypertension: a review of diagnosis and management. JAMA. 2014;311(21):2216–24.
Bradford JR. The innervation of the renal blood vessels. J Physiol. 1889;10(5):358–432.
Adson A, Craig W, Brown G. Surgery in its relation to hypertension. Surg Gynecol Obstet. 1936;62:314–31.
Papademetriou V, Rashidi AA, Tsioufis C, Doumas M. Renal nerve ablation for resistant hypertension: how did we get here, present status, and future directions. Circulation. 2014;129(13):1440–51.
Smithwick RH, Thompson JE. Splanchnicectomy for essential hypertension; results in 1,266 cases. J Am Med Assoc. 1953;152(16):1501–4.
Peet MM, Woods WW, Braden S. The surgical treatment of hypertension: results in 350 consecutive cases treated by bilateral supradiaphragmatic splanchnicectomy and lower dorsal sympathetic ganglionectomy. J Am Med Assoc. 1940;115(22):1875–85.
Freis ED, Wanko A, Wilson IM, Parrish AE. Treatment of essential hypertension with chlorothiazide (Diuril): its use alone and combined with other antihypertensive agents. J Am Med Assoc. 1958;166(2):137–40.
DiBona GF, Kopp UC. Neural control of renal function. Physiol Rev. 1997;77(1):75–197.
Persu A, Maes F, Renkin J, Pathak A. Renal denervation in hypertensive patients: back to anatomy? Hypertension. 2020;76(4):1084–6.
Ram CVS. Status of renal denervation therapy for hypertension. Circulation. 2019;139(5):601–3.
Townsend RR, Mahfoud F, Kandzari DE, Kario K, Pocock S, Weber MA, Ewen S, Tsioufis K, Tousoulis D, Sharp ASP, et al. Catheter-based renal denervation in patients with uncontrolled hypertension in the absence of antihypertensive medications (SPYRAL HTN-OFF MED): a randomised, sham-controlled, proof-of-concept trial. Lancet. 2017;390(10108):2160–70.
Sakakura K, Ladich E, Cheng Q, Otsuka F, Yahagi K, Fowler DR, Kolodgie FD, Virmani R, Joner M. Anatomic assessment of sympathetic peri-arterial renal nerves in man. J Am Coll Cardiol. 2014;64(7):635–43.
Page IH, Heuer GJ. The effect of renal denervation on patients suffering from nephritis. J Clin Invest. 1935;14(4):443–58.
Page IH, Heuer GJ. The effect of renal denervation on the level of arterial blood pressure and renal function in essential hypertension. J Clin Invest. 1935;14(1):27–30.
Bhatt DL, Kandzari DE, O’Neill WW, D’Agostino R, Flack JM, Katzen BT, Leon MB, Liu M, Mauri L, Negoita M, et al. A controlled trial of renal denervation for resistant hypertension. N Engl J Med. 2014;370(15):1393–401.
Patel HC, Hayward C, Vassiliou V, Patel K, Howard JP, Di Mario C. Renal denervation for the management of resistant hypertension. Integr Blood Press Control. 2015;8:57–69.
Haemmerich D, Schutt DJ. RF ablation at low frequencies for targeted tumor heating: in vitro and computational modeling results. IEEE Trans Biomed Eng. 2010;58(2):404–10.
Tungjitkusolmun S, Staelin ST, Haemmerich D, Tsai JZ, Webster JG, Lee FT Jr, Mahvi DM, Vorperian VR. Three-dimensional finite-element analyses for radio-frequency hepatic tumor ablation. IEEE Trans Biomed Eng. 2002;49(1):3–9.
Tungjitkusolmun S, Woo EJ, Cao H, Tsai JZ, Vorperian VR, Webster JG. Finite element analyses of uniform current density electrodes for radio-frequency cardiac ablation. IEEE Trans Biomed Eng. 2000;47(1):32–40.
Leertouwer TC, Gussenhoven EJ, van Jaarsveld BC, van Overhagen H, Bom N. Man in ’t Veld AJ: in-vitro validation, with histology, of intravascular ultrasound in renal arteries. J Hypertens. 1999;17(2):271–7.
Rossmann C, Haemmerich D: Review of temperature dependence of thermal properties, dielectric properties, and perfusion of biological tissues at hyperthermic and ablation temperatures. Crit Rev™ Biomed Eng. 2014; 42(6):467–92
Zurbuchen U, Holmer C, Lehmann KS, Stein T, Roggan A, Seifarth C, Buhr HJ, Ritz JP. Determination of the temperature-dependent electric conductivity of liver tissue ex vivo and in vivo: Importance for therapy planning for the radiofrequency ablation of liver tumours. Int J Hyperthermia. 2010;26(1):26–33.
Berjano EJ. Theoretical modeling for radiofrequency ablation: state-of-the-art and challenges for the future. Biomed Eng Online. 2006;5(1):24.
Mikhail AS, Negussie AH, Graham C, Mathew M, Wood BJ, Partanen A. Evaluation of a tissue-mimicking thermochromic phantom for radiofrequency ablation. Med Phys. 2016;43(7):4304.
Greene ER, Venters MD, Avasthi PS, Conn RL, Jahnke RW. Noninvasive characterization of renal artery blood flow. Kidney Int. 1981;20(4):523–9.
Sata Y, Head GA, Denton K, May CN, Schlaich MP. Role of the sympathetic nervous system and its modulation in renal hypertension. Front Med (Lausanne). 2018;5:82.
Rasband WS, ImageJ, U. S. National Institutes of Health, Bethesda, Maryland, USA, https://imagej.nih.gov/ij/, 1997–2018.
Acknowledgements
Not applicable.
Authors’ information
Not applicable.
Funding
This work was funded in part by Eng. Thomas Osypka, CEO and President of Oscor, Inc., Palm Harbor, Florida, USA.
Author information
Authors and Affiliations
Contributions
NP conducted the experiments and wrote the manuscript; KM and SES advised on the experiments' design. All authors reviewed the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Not applicable.
Competing interests
Noel Perez was previously the Vice President of Research and Development at Oscor, Inc., who sponsored this work.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Pérez, N., Muffly, K. & Saddow, S.E. Technical advance in silico and in vitro development of a new bipolar radiofrequency ablation device for renal denervation. BMC Cardiovasc Disord 21, 500 (2021). https://doi.org/10.1186/s12872-021-02305-x
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12872-021-02305-x