EnSite NavX mapping system guided implantation of a dual-chamber permanent pacemaker in a 41-year-old pregnant woman with a 4-year follow-up

Abstract

Background

X-ray fluoroscopy has been the primary cardiac imaging modality in permanent pacemaker implantation (PPI) operations, but it inevitably results in radiation exposure for both operators and patients. Fluoroscopy is considered a contraindication, especially in certain circumstances, such as gestation, during which the fetus is most sensitive to radiation exposure. Therefore, measures to avoid radiation exposure are necessary, and a more safe and feasible approach is needed for this procedure. Since the EnSite NavX mapping system (ENMS) can create the required geometric contours of those relevant cardiac structures and chambers, it can be used as an alternative to X-ray fluoroscopy in PPI. In addition, because the displacement of atrial leads is a common complication of PPI, lead displacement may occur more readily without fluoroscopic guidance. Therefore, reliable measures are required to prevent leads from displacement.

Case introduction

A 41-year-old woman at the 15th week of gestation was referred to our department with recurrent episodes of syncope and amaurosis fugax for 2 years. Holter monitoring showed sinus rhythm, Mobitz Type II atrioventricular block and high-grade atrioventricular block with ventricular arrest up to 4945 ms. A dual-chamber PPI was performed successfully for the patient under the guidance of the ENMS instead of fluoroscopy. Displacement of atrial lead was effectively avoided by bending the top of atrial lead before implantation and making it a U-shape during operation, which left space for possible subsequent external pulling stress.

Conclusions

For PPI, ENMS is a feasible and reliable alternative to traditional X-ray fluoroscopy, especially when performing operations on pregnant patients. By bending the top of the active-fixation atrial lead into a U-shape during operation, the displacement of atrial lead may be avoided.

Peer Review reports

Background

X-ray fluoroscopy is routinely used for cardiac imaging in permanent pacemaker implantation (PPI) operations. However, fluoroscopy is considered a contraindication that should be avoided in some special cases, such as pregnant women and children. Therefore, other reliable alternatives are desired. As the EnSite NavX mapping system (ENMS) can create the required geometric contours of those relevant cardiac structures and chambers, it can be used as a replacement for X-ray fluoroscopy in PPI [1]. Additionally, since the displacement of atrial leads is a common complication after PPI operations, lead displacement may tend to occur more easily without fluoroscopic guidance. Thus, reliable measures are required to prevent leads from displacement. This report presents a successful operation of PPI guided by ENMS for a 41-year-old pregnant woman with 4-month-old twins. Effective techniques are applied to prevent the atrial lead from displacement in this case.

Case presentation

A 41-year-old pregnant woman at the 15th week of gestation with recurrent episodes of syncope and amaurosis fugax for 2 years was referred to our department on July 16, 2017. Physical examination: blood pressure 118/85 mmHg; heart rate 58 bpm, irregular, no cardiac murmurs. The 24-h Holter monitoring showed sinus rhythm, Mobitz Type II atrioventricular block (Fig. 1, left), high-grade atrioventricular block with ventricular arrest up to 4945 ms (Fig. 1, right) when the patient suffered from amaurosis fugax. Echocardiographic examination showed negative. Serum creatine kinase (CK) 33u/l, creatine kinase MB subfraction (CK-MB) 11u/l, and other laboratory results were also negative. The patient had no hypertension, cardiomyopathy or other cardiovascular diseases. She had no any family history of genetic diseases either. Therefore, acute myocardial injury was excluded as the cause of the atrioventricular block. Finally, a dual-chamber PPI was performed under the guidance of ENMS to prevent the pregnant woman and the fetus from radiation exposure.

Fig. 1
figure 1

A 24-h Holter monitoring shows Mobitz Type II atrioventricular block (left) and high-grade atrioventricular block with ventricular arrest up to 4945 ms (right) associated with syncope

The operation was performed on July 18, 2017. The method of ENMS for drawing the three-dimensional anatomical map of heart chambers was identical to that in the radiofrequency ablation operation. First, three pairs of mutually perpendicular electrode pads were attached around the patient's heart to form a stable special electric field. In this way, when a metal electrode catheter with microcurrent connected to ENMS was placed in this electric field, its precise three-dimensional position could be displayed on the screen. Then a 7F sheath was inserted into the right femoral vein under local anesthesia. After connecting an alligator cable to the ENMS, a 10-polar catheter was inserted into the sheath and advanced to the inferior and superior vena cava, right atrium, tricuspid annulus, and right ventricle to create the geometry contours of those chambers by wagging and moving it inside those cavities (Fig. 2). Another 7F sheath was inserted into the left subclavian vein through a percutaneous puncture. Under the guidance of ENMS, the active-fixation leads of the right atrium and right ventricle were manipulated and secured to proper positions (Fig. 3). Based on our previous experience, to avoid the displacement of atrial leads, the front of the atrial lead was bent into a curved shape before implantation (Fig. 4, left). The atrial lead was continued to be pushed forward by about 5 cm after implantation in the proper position under transthoracic echocardiography guidance to make the front of the atrial lead a U-shape. The parameters were tested, as shown in Table 1. Then the leads were connected to the pacemaker in DDD mode, and the pacemaker was implanted into the left infraclavicular subcutaneous pocket. The total operation time was 68 min. The patient was discharged and followed up regularly. Boy and girl twins were successfully delivered by caesarean section 4 months after the pacemaker operation. In the 5th month after the operation, though the curvature of atrial leads became smaller than its original curvature (Fig. 4, right) and the sensed amplitude of atrial leads decreased, other parameters, such as the pacing thresholds and impedance of the two leads, were normal (Table 1). The patient had no relevant unpleasant symptoms. The two leads were regularly tested for 4 years without any intervention.

Fig. 2
figure 2

The three-dimensional anatomical map of heart chambers drawn by ENMS. SVC, superior vena cava; IVC, inferior vena cava; RA, right atrium; RV, right ventricle; CS, coronary sinus

Fig. 3
figure 3

The right anterior oblique view of the patient shows the positions of the two leads. SVC, superior vena cava; IVC, inferior vena cava; RA, right atrium; RV, right ventricle

Fig. 4
figure 4

Atrial lead with a bigger bend before implantation (left). X-ray shows the two positions of leads 5 months after cesarean section (right)

Table 1 Lead parameters in different periods

Discussion

Fluoroscopy is the primary cardiac imaging method in PPI operations, but it inevitably exposes operators and patients to radiation. Fluoroscopy is considered a contraindication for pregnant women, especially those between the 8th to 15th week of pregnancy, when the fetus is most sensitive to radiation. Thus, a more feasible cardiac imaging approach is needed instead of fluoroscopy. Currently, in international medical guidelines, no methodological recommendations are available on how to avoid or reduce radiation exposure for PPI in pregnant women. In some reported cases [1, 2], transthoracic or transesophageal or even intracardiac echocardiography was used to guide the implantation of pacemaker leads. However, ultrasound is not widely used for cardiac imaging in PPI due to the visualization problems of all leads or complications such as aspiration pneumonia or suffocation [3].

The ENMS is based on the measurement of electrical currents for the three-dimensional localization of catheters. A low-amplitude, high-frequency electrical current was delivered through three pairs of skin patches located on the patient's chest in orthogonal planes, creating a voltage gradient along each direction and acquiring geometric contours of those relevant cardiac structures. At present, ENMS has been used in radiofrequency ablation treating cardiac arrhythmia, such as atrial fibrillation and premature ventricular contraction. Ping Guo et al. [3] reported six cases of PPI with zero-fluoroscopy under the guidance of ENMS, including one pregnant woman. No complications were observed during 6 months of follow-up. An Italy multicentre registry study during 2011–2014 [4] on cardiac resynchronization therapy (CRT) device implantation concluded that ENMS-guided CRT implantation was proven safe and effective in both high- and low-experienced centers and could reduce fluoroscopy. Clinical studies suggested that ENMS may become the trend for cardiac imaging to guide PPI instead of fluoroscopy to reduce radiation exposure.

Displacement of atrial leads is a common complication after PPI operations, and late displacement is defined as the displacement occurring more than 6 weeks after pacing system implantation [5, 6]. Generally, after PPI, the upper body and respiratory movement of the patient will inevitably cause a certain degree of pulling force on the pacing leads, which may lead to lead displacement and a significant increase in the pacing threshold. In particular, displacement of pacing leads during delivery could be caused by many factors, such as deep breathing in the follow-up delivery process, large-scale changes in body posture, cesarean section, and mechanical ventilation used for anesthesia. Therefore, an effective technique was adopted in this case to prevent lead displacement. In this way, some space could be left for possible external pulling stress after the operation. Even if the U-shaped arc was straightened, the tip of atrial lead would not be dislocated, thereby ensuring the stability of the pacing threshold. According to the post operative follow-up, this technique successfully prevented lead from displacement. The threshold value did not change in the follow-up after the cesarean section. However, the sensing P-wave lightly decreased, possibly because the U-shaped arc of the atrial lead was straightened to some extent, which was confirmed by subsequent X-ray fluoroscopy examination. As the lead tip did not move, the stability of other parameters, such as pacing threshold and impedance, and the normal pacing function were guaranteed.

Conclusion

This study revealed that ENMS is a feasible and reliable alternative to traditional X-ray fluoroscopy in PPI operations, especially performing such operations on pregnant patients. Displacement of atrial lead can be avoided by bending the front part of the active-fixation lead of the right atrium into a U-shape in PPI operation.

Availability of data and materials

The datasets used and analyzed in the present report are available from the corresponding author on reasonable request.

Abbreviations

PPI:

Permanent pacemaker implantation

ENMS:

EnSite NavX mapping system

LVEF:

Left ventricular ejection fraction

CRT:

Resynchronization therapy

References

  1. Gianni C, Domenico G, Della R, Natale A. Fluoroless 3D mapping-guided pacemaker implant in a pregnant patient. Pacing Clin Electrophysiol. 2021;44:1641–5.

    Article  Google Scholar 

  2. Antonelli D, Bloch L, Rosenfeld T. Implantation of permanent dual chamber pacemaker in a pregnant woman by transesophageal echocardiographic guidance. PACE. 1999;22:534–5.

    Article  CAS  Google Scholar 

  3. Guo P, Qiu J, Wang Y, et al. Zero-fluoroscopy permanent pacemaker implantation using Ensite NavX system: clinical viability or fanciful technique. Pacing Clin Electrophysiol. 2018;41:122–7.

    Article  Google Scholar 

  4. Del Greco M, Maines M, Marini M, et al. Three-dimensional electroanatomic mapping system-enhanced cardiac resynchronization therapy device implantation: results from a multicenter registry. J Cardiovasc Electrophysiol. 2017;28(1):85–93.

    Article  Google Scholar 

  5. Guan Fu, et al. Delayed management of atrial lead displacement after pacemaker implantation: a case report. J Med Case Rep. 2021;15(1):9–9.

    Article  Google Scholar 

  6. Chauhan A, Grace AA, Newell SA, Stone DL, Shapiro LM, Schofeld PM, Petch MC. Early complications after dual chamber versus single chamber pacemaker implantation. Pacing Clin Electrophysiol. 1994;17(11):2012–5.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

We would like to express our gratitude to Li LI, who helped us perform an ultrasonic examination during PPI.

Funding

This work was supported by the Teaching Research Project of Air Force Medical Center (2021JX001) and the Military Logistical Special Project for Health Care (18BJZ07). The funding body had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

JHW and YJC contributed to the conception and design of the report. Material preparation, data collection and analysis were performed by WWZ, HS, ZYZ, LA and MW. The first draft of the manuscript was written by PW and GSW. All authors commented on previous versions of the manuscript and have read and approved the final manuscript.

Corresponding author

Correspondence to Yan-Jie Cao.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Institutional Ethics Committee of the Air Force Medical Center, and written informed consent was also obtained from the participant. The study adhered to the tenets of the Declaration of Helsinki.

Consent for publication

Not applicable.

Competing interests

All authors of this publication have no conflict of interest to declare. The current manuscript or part of it has not been published before.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wang, P., Wei, GS., Wang, JH. et al. EnSite NavX mapping system guided implantation of a dual-chamber permanent pacemaker in a 41-year-old pregnant woman with a 4-year follow-up. BMC Cardiovasc Disord 22, 325 (2022). https://doi.org/10.1186/s12872-022-02764-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12872-022-02764-w

Keywords