- Research article
- Open access
- Published:
Impact of chronic outward force on arterial responses of proximal and distal of long superficial femoral artery stent
BMC Cardiovascular Disorders volume 21, Article number: 323 (2021)
Abstract
Background
Self-expanding nitinol stent (SENS) implantation is commonly oversized in the superficial femoral artery (SFA), and leads to chronic outward force (COF) and in-stent restenosis (ISR). This study aimed to investigate the impact of COF of oversizing SENS on ISR of SFA.
Methods
In patients with implanted SENS in SFA, intimal hyperplasia especially between proximal segment and distal segment was evaluated by quantitative angiography, and the impact of COF on mid-term angiographic outcomes was investigated. In addition, porcine model with implanted SENS was used to evaluate the impact of COF on angiographic and histopathologic outcomes at 1 month. Excised stented arteries were evaluated by histopathologic analysis.
Results
We analyzed 65 SENS in 61 patients with follow-up angiography at 6 months to 1 year. The baseline diameter was 6.8 ± 0.71 mm and length were 97.0 ± 33.8 mm for the SENS. The ratio of the diameter of the stent to the reference vessel was 1.3 ± 0.24 at the proximal portion and 1.53 ± 0.27 at the distal portion (P < 0.001). In the long SFA stent, stent-to-vessel ratio was significantly higher in the distal stent than in the proximal stent (1.3 ± 0.2 vs. 1.55 ± 0.25, P = 0.001). ISR incidence was higher at the distal stent (37.3% vs 52.6%, P = 0.029). All 11 pigs survived for 4 weeks after SENS implantation. The vessel diameter was 4.04 ± 0.40 mm (control group) vs 4.45 ± 0.63 mm (oversized group), and the implanted stent diameter was 5.27 ± 0.46 mm vs. 7.18 ± 0.4 mm (P = 0.001). The stent-to-vessel diameter ratio was 1.31 ± 0.12 versus 1.63 ± 0.20 (P < 0.001). After 4 weeks, restenosis % was 29.5 ± 12.9% versus 46.8 ± 21.5% (P = 0.016). The neointimal area was 5.37 ± 1.15 mm2 vs. 8.53 ± 5.18 mm2 (P = 0.05). The restenosis % was 39.34 ± 8.53% versus 63.97 ± 17.1% (P = 0.001).
Conclusions
COF is an important cause of restenosis in the distal portion of the SFA stent. Optimal sizing of the SFA stent is important to reduce the incidence of restenosis. Therefore, COF was an important factor of restenosis following distal SFA stenting.
Introduction
Balloon angioplasty and stent implantation are popular in patients with peripheral artery disease (PAD) [1,2,3]. The self-expanding nitinol stent (SENS) for superficial femoral artery (SFA) lesions is morphologically and clinically superior to balloon angioplasty [2, 4, 5]. However, stent size influences clinical outcomes [6, 7]. Implanted SENS is commonly oversized in the SFA, especially at the distal portion due to tapered diameter of SFA for long SENS [8,9,10]. SFA stents are under dynamic stress to external forces including bending, twisting, torsion, elongation, foreshortening and external compression [11, 12]. The mechanical stresses to the SFA are associated with an increased risk of stent fracture and subsequent significant restenosis [11, 13]. The chronic outward force (COF) exerted by a SENS is another important factor to cause restenosis, and several investigations evaluated the effect of SENS oversizing on the risk of in-stent restenosis (ISR) [6, 14,15,16]. COF could produce continuous mechanical stimulation of the arterial wall, which can increase neointimal hyperplasia and lead to significant ISR [17].
This study aimed to evaluate the impact of SENS oversizing on ISR of SFA. We examined clinical and animal experiments data to determine the impact of COF on the arterial wall, and investigated the impact of SENS oversizing with respect to continued expansion following stenting and the mid-term histological impact of oversized stent on SFA.
Methods
Study design
The clinical evaluation focused on intimal hyperplasia especially between the proximal and distal segments by implanted SENS in SFA. Quantitative angiography (QA) was used to determine the ISR rate following mid-term catheter angiography. The impact of COF on mid-term angiographic and clinical outcomes was investigated (Fig. 1).
Preclinical data were obtained from a porcine model following SENS (CV Bio, Seoul, Korea) implantation. The model was used to evaluate the impact of COF on angiographic outcomes and histopathologic outcomes at 1 month. Control pigs received an adequately sized SENS, whereas the COF group received an over-sized stent 1.5 to 1.8 times larger than the control size. At 4 weeks, catheter angiography and histopathologic evaluation was performed (Fig. 1).
Patients
The effect of COF was explored by a retrospective review of the medical records of 329 patients who underwent SENS implantation for SFA from January 2004 to June 2016. The data were obtained from the percutaneous transluminal angioplasty (PTA) registry of Korea University Guro Hospital (KUGH) (Seoul, South Korea). PTA registry is a single-center, prospective, all-comer registry designed to reflect “real world” practice since 2004. Patients had signs of a moderate intermittent claudication (Rutherford stage 2), severe intermittent claudication (Rutherford stage 3) or ischemic pain (Rutherford stage 4), or significant atherosclerotic effect in one or both femoral arteries (stenosis diameter > 70%).
Data were collected by a trained study-coordinator, following a standardized case report form. The participants or their legal guardians provided written consent to participate in the study. The study protocol was approved by Medical Review Board of KUGH.
Percutaneous intervention
Aspirin and clopidogrel were used in all patients and 70 U/kg heparin were given after the start of the procedure. After local anesthesia with Lidocain 2% 5F or 6F sheath was delivered through the femoral artery by the Seldinger approach. If balloon angioplasty or stenting proved technically feasible by ipsilateral antegrade approach, an antegrade puncture was preferred. Otherwise, a contralateral femoral approach or retrograde puncture from the distal SAF or proximal tibial access for bi-directional approach was used.
A 5-Fr Omni internal mammary catheter, or Judkin's right catheter was inserted into the aortic-iliac artery branch when contralateral approach was needed. The introducing soft Terumo wire was inserted into the lesion, and then the guiding sheath was placed for SFA treatment. Most frequently used guiding sheaths were 6F Ansel Checkflo (Cook, USA), 6-8F Balkin sheath (Cook, Bloomington, USA) or 8F Contralateral-1 guiding sheath (Cordis, USA).
After the successful wire crossing by 018 or 035 long wires, conventional balloon angioplasty was performed. The balloon dilatation duration was extended for at least 120 s with adequately sized balloon (4.0–7.0 mm diameter with longer balloon). Once the balloon response was not favorable, including flow limiting intimal dissection, significant residual stenosis or thrombus formation, SENS was implanted. Angiographic success was considered to be the case when residual stenosis was < 30% after the procedure and blood flow was well maintained through and distal to the target lesion.
Clinical and angiographic follow-up
Patients with successful intervention were clinically followed at 30 days and 3 months respectively to monitor the diabetic foot recovery or the severity of symptoms. Patients were strongly recommended for mid-term imaging follow-up by invasive or CT angiography at 6 to 12 months. QA was used to assess the impact of COF on ISR in proximal and distal SFA stents.
Pig COF model and procedures
Animal experiments were conducted at the K-BIO Osong Animal Research Center and the protocols were approved by Animal Care and Use Review Board (No. KBIO-IACUC-2016–005) according to the guidelines by the National Institute of Health. Eleven non-atherosclerotic female pigs (weight 25–35 kg) were assigned. Twenty self-expanding nitinol stents (SENS-CV Bio, Seoul, Korea) 5.0–8.0 mm in diameter and 30–50 mm in length were used.
Aspirin (100 mg/day) and clopidogrel (75 mg/day) were given 3 days before the procedure. On the day of the procedure, after general anesthesia with isoflurane followed by respiratory anesthesia using an animal ventilator, the right carotid artery was incised, the carotid artery was exposed, a 6-F sheath was inserted, and 200 IU/kg of heparin was injected. The 6-F guiding catheter was inserted into the SFA and baseline SFA angiography was performed. SFA in the control group was treated using SENS of the appropriate size. SFA in the COF group was treated by an oversized stent. After the procedure, the carotid artery was ligated, the pigs were transferred to the breeding farm and observed for 4 weeks. During this period, aspirin and clopidogrel were administered continuously.
The effect of COF was analyzed by angiography and histopathologic results. After 4 weeks, catheter angiography was performed through the opposite carotid artery, and QA using CAAS software was used for the evaluation of stent patency, fracture, and ISR degree. To be euthanized, pigs are given an intravenous carbon dioxide/oxygen gas mixture after general anesthesia with isoflurane and excised stented arteries were fixed in 10% formalin for 24 h for histopathologic analysis of inflammation, necrosis and intimal hyperplasia.
Histomorphometry
The tissues were embedded in paraffin and cut into fine section (50–100 μm) using a low-speed diamond wafer mounted on an isoment saw (Buehler Ltd., Lake Bluff, IL, USA). After hematoxylin and eosin (H&E) staining, the sections were analyzed by calibrated microscope digital video imaging system and a VISUS 2000 computerized visual image analysis system. The external elastic lamina (EEL) area, internal elastic lamina (IEL) area and the lumen area of the central, proximal and distal part of the stent from pig SFA were examined. The area stenosis of the stent vessels was conventionally calculated as: 100 × (1-lm area/IEL area).
The injury score of the vessel wall by the stent and inflammation score were evaluated by Schwartz method [18]. The injury score of the vessel wall was scored as 0 for no endothelial cell damage, 1 for the break in the internal elastic membrane, 2 for the media membrane injury, and 3 for the damage of the external elastic membrane to the adventitia. The inflammation score was scored as 0 for inflammatory cells did not infiltrate around the strut, 1 for the cells did not surround the strut, 2 for the cells did not surround the strut and 3 for tight densification.
Statistical analysis
All statistical analyses were performed using SPSS for Windows version 20.0 (SPSS Inc., Chicago, USA). The normality was verified via one-sample Kolmogorov–Smirnov test. Continuous variables are expressed as mean ± standard deviation (SD), categorical data were reported as frequencies and percentages. Unpaired Student's t-test were used to compare continuous variables while Chi-square test was used to compare categorical variables. Correlation between the stent/vessel ratio and the stenosis results was determined using linear regression analysis. The receiver operating characteristic (ROC) analysis was performed to get area under the curve (AUC). P < 0.05 was considered statistically significant.
Results
Clinical results
Sixty-five SENS of SFA in 61 patients with follow-up angiography at 6 months to 1 year were analyzed. The mean follow-up duration was 302 days. All PTA procedures and follow-up angiography were successful. In the most severe lesions, the minimal luminal diameter (MLD) was 0.85 ± 1.32 at the proximal lesion site mm and 0.91 ± 1.15 mm at distal lesion site (P = 0.619). The mean stent diameter and length was 6.8 ± 0.71 mm and 97.0 ± 33.8 mm, respectively. The ratio of the diameter of the stent to the reference vessel was 1.3 ± 0.24 at the proximal portion and 1.53 ± 0.27 at the distal portion (P < 0.01) (Table 1). QA of the follow-up angiograms showed a trend toward higher incidence of ISR in the distal portion of the stent. However, the mean percent restenosis was significantly higher and mean MLD was lower in the distal portion than in the proximal portion (Table 2). Subgroup analysis according to the stent length showed that for stents longer than 100 mm, the stent-to-vessel ratio and the incidence of ISR was significantly higher in the distal site than in the proximal site. The mean percent restenosis was also significantly higher in the distal site (Table 3). In contrast, for the stents shorter than 100 mm, there were no significant difference in the incidence of ISR and mean MLD between the two groups, except a trend toward higher mean percent stenosis (Table 4).
Animal study results
All 11 animals survived for 4 weeks without acute limb ischemia due to stent thrombosis. After the procedure, quantitative and morphological analyses were performed (Fig. 1). Before stenting, the vessel diameter was 4.04 ± 0.40 in the control group and 4.45 ± 0.63 in the oversized group (P = 0.12). The mean stent diameter implanted in the control group was 5.27 ± 0.46 mm and 7.18 ± 0.4 mm in the oversized group (P = 0.001). The stent-to-vessel diameter ratio was 1.31 ± 0.12 versus 1.63 ± 0.20 in control and oversized groups, respectively (P < 0.001) (Table 5). After 4 weeks, the incidence of ISR was higher, MLD was smaller, and the mean percentage of restenosis was higher in oversized group (Table 6). The diagnostic criteria for stent/artery ratios were 1.51 times of the standard for diagnosing the stenosis using ROC analysis (sensitivity of 0.791 and specificity of 0.929). Correlation between oversizing ratio was statistically significant (R = 0.803, P = 0.002). In the histopathologic examination, the incidence of ISR was higher, lumen area was smaller, neointimal and stenosis area were larger in oversized group. There was a trend toward higher inflammation score in oversized group (Table 7). In addition, the correlation of stenosis area and the artery-to-stent oversized ratio was significant (R = 0.783, P = 0.010) (Fig. 2).
Discussion
The major findings of the impact of COF on arterial responses of SFA stent are summarized as follows. First, clinical study showed that for long SFA stent, the stent-to-artery ratio was significantly higher in the distal stent than in the proximal stent. The incidence of ISR and mean percentage of restenosis was significantly higher at the distal site, particularly in the implanted stent longer than 100 mm. Second, in animal models, the incidence of ISR was higher, lumen area was smaller, and neointimal and stenosis area were larger and there was a trend toward higher inflammation score in the oversized group. Third, the stent to artery ratio was 1.51 times of the standard for diagnosing stenosis using ROC analysis. These results support the idea that stent oversizing significantly increases the intramural stress to stimulate smooth muscle in the vessel, leading to inflammatory response and pathologic stress in the artery wall that cause irreversible damage [19].
Several clinical trials demonstrated the benefit of SENS implantation over angioplasty for stenosis or occlusion of the SFA [2, 4, 20]. However, SFA poses unique challenge for endovascular stenting. SFA is a long muscular artery that is fixed between the hip and the knee. Complex motions of the hip joint can impose complex external mechanical stresses on the SFA, including flexion, compression and torsion [11]. SFA is further exposed to longitudinal and lateral compressional stress as the artery dives through the Hunter canal between the muscle bodies of the anterior and medial compartments of the thigh [12]. Currently, a stent diameter of 1–2 mm oversizing is recommended in SFA stent implantation [8, 21]. For a self-expanding non-tapered nitinol stent, it will be implanted in accordance with the proximal diameter of the reference vessel, since the stent is self-tapered at the distal site after stenting. Therefore, the distal portion of the long SENS will exert stronger COF to distal SFA, leading to higher incidence of intimal hyperplasia and subsequent restenosis.
Previous studies reported that average reference vessel diameter of SFA is 5 mm, and most commercially available stent has a diameter of 7 mm [8,9,10, 22, 23]. However, SFA consists of tapered vessels. Therefore, in this procedure, a larger diameter stent will be implanted into the smaller SFA distal portion. The different diameters of the proximal and distal portions of stent impose a wall stress on the arterial wall. Vessel wall injury is an important factor during stent implantation [15]. The main reason for restenosis is adverse remodeling caused by endothelial injury, inflammation of atherosclerosis plaque, hemodynamic factors, and mechanical stress from a permanent stent [24].
The incidence of ISR depends on the type, design and length of the implanted stent. Innovative stent design may decrease ISR and improve clinical outcomes. The unique design of the Supera Peripheral Stent (Abbott Vascular, Abbott Park, IL) aimed to overcome issues related to areas of high flexion and minimal COF [9]. Recently, the Supera Interwoven Nitinol Stent Outcomes in the Above-Knee Interventions (SAKE) study showed high rate of patency at 6 and 12 months with no stent restenosis for femoropopliteal disease treated with novel interwoven-wire Supera stent [25]. Furthermore, the outcomes after popliteal stenting with the Supera stent from the Leipzing Supera popliteal artery registry showed 6- and 12-month primary patency rate of 94.6% and 87.7%, respectively, with 4 patients experiencing in-stent occlusions and 6 patients experiencing in-stent restenosis [26]. Optimizing stent design can improve stent performance [27].
The length of the SENS is also an important factor of ISR. Long stenosis (> 100 mm) had poor outcome, with restenosis rate exceeding 70% at 1 year [28]. The restenosis rates vary between 35 and 65% at 6–12 months for lesions > 100 cm [29]. In our study, the restenosis rate was significantly higher in the distal portion than in the proximal portion of the long stent (> 100 mm). Conversely, there was no difference in short lesions (< 100 mm). These results revealed the importance of COF caused by oversized SENS in the distal SFA.
Stents are designed to be larger in diameter than the healthy artery, which is referred as “stent oversizing” [19]. The significant diameter change usually occurs in carotid artery stenting, but also occurs in femoral arteries [19]. The mechanical environment of peripheral arteries could be the predominant cause of relatively higher restenosis rate [13]. Although the ideal oversizing ratio is stent specific, nitinol stent oversizing has a very small impact on the immediate lumen gain [30]. However, severe oversizing (stent-to-artery ratio > 1.4:1) resulted in profound long-term histological changes including exuberant neointimal proliferation and luminal stenosis in animal models [6]. These results suggest that the oversizing ratio is an important factor for neointimal formation in SENS, and self-expanding stents without exceeding an oversizing ratio of 1.4 is crucial for long-term patency in peripheral artery stenting [16]. Furthermore, when a non-tapered stent is placed in vessels with a large discrepancy in diameter, attention must be paid for increased neointimal hyperplasia in the oversized side [17]. In our porcine model, the cutoff ratio of SENS to artery ratio was calculated to be 1.51. The restenosis rate was significantly higher in oversized group than in control group.
In summary, we investigated the impact of oversized SENS implanted in tapered SFA arteries on restenosis caused by COF. Oversizing of the stent is an important determinant of final arterial wall diameter and stress. Therefore, SENS with optimal oversizing should be designed in cases with significant arterial tapering.
Study limitations
In clinical study, our single-center retrospective study design only allowed for analysis based on our routine clinical practice. The sample size did not provide sufficient power to detect ISR incidence or compare clinical events. Operators were not blinded to baseline measures. Finally, our data reflected in-hospital outcomes, and we did not analyze long-term outcomes. Therefore, future randomized studies are needed to confirm our conclusions.
In animal study, the sample size of 22 stents was small, and we did not find significant differences in incidence of restenosis or related parameters between the two groups. In addition, we used healthy animals with non-atherosclerotic arteries. Although blood vessels of pig are very similar to those of human, the results of animal experiments may not be same as those of clinical studies.
Conclusions
COF is an important cause of restenosis in the distal portion of the SFA stent. Optimal sizing of the SFA stent implantation with SENS is important to reduce the incidence of restenosis. Therefore, COF was an important factor of restenosis following distal SFA stenting.
Availability of data and materials
The datasets supporting the conclusions of this article are included within the article.
References
Norgren L, Hiatt WR, Dormandy JA, Nehler MR, Harris KA, Fowkes FG, Group TIW. Inter-society consensus for the management of peripheral arterial disease (TASC II). J Vasc Surg 2007;45(Suppl 5): S5–67.
Laird JR, Katzen BT, Scheinert D, Lammer J, Carpenter J, Buchbinder M, Dave R, Ansel G, Lansky A, Cristea E, Collins TJ, Goldstein J, Jaff MR, Investigators R. Nitinol stent implantation versus balloon angioplasty for lesions in the superficial femoral artery and proximal popliteal artery: twelve-month results from the RESILIENT randomized trial. Circ Cardiovasc Interv. 2010;3:267–76.
Schillinger M, Sabeti S, Dick P, Amighi J, Mlekusch W, Schlager O, Loewe C, Cejna M, Lammer J, Minar E. Sustained benefit at 2 years of primary femoropopliteal stenting compared with balloon angioplasty with optional stenting. Circulation. 2007;115:2745–9.
Schillinger M, Sabeti S, Loewe C, Dick P, Amighi J, Mlekusch W, Schlager O, Cejna M, Lammer J, Minar E. Balloon angioplasty versus implantation of nitinol stents in the superficial femoral artery. N Engl J Med. 2006;354:1879–88.
Krankenberg H, Schluter M, Steinkamp HJ, Burgelin K, Scheinert D, Schulte KL, Minar E, Peeters P, Bosiers M, Tepe G, Reimers B, Mahler F, Tubler T, Zeller T. Nitinol stent implantation versus percutaneous transluminal angioplasty in superficial femoral artery lesions up to 10 cm in length: the femoral artery stenting trial (FAST). Circulation. 2007;116:285–92.
Zhao HQ, Nikanorov A, Virmani R, Jones R, Pacheco E, Schwartz LB. Late stent expansion and neointimal proliferation of oversized Nitinol stents in peripheral arteries. Cardiovasc Intervent Radiol. 2009;32:720–6.
Stoeckel D, Pelton A, Duerig T. Self-expanding nitinol stents: material and design considerations. Eur Radiol. 2004;14:292–301.
Gray WA, Feiring A, Cioppi M, Hibbard R, Gray B, Khatib Y, Jessup D, Bachinsky W, Rivera E, Tauth J, Patarca R, Massaro J, Stoll HP, Jaff MR, S.S. Investigators. S.M.A.R.T. self-expanding nitinol stent for the treatment of atherosclerotic lesions in the superficial femoral artery (STROLL): 1-year outcomes. J Vasc Interv Radiol 2015;26:21–28.
Garcia L, Jaff MR, Metzger C, Sedillo G, Pershad A, Zidar F, Patlola R, Wilkins RG, Espinoza A, Iskander A, Khammar GS, Khatib Y, Beasley R, Makam S, Kovach R, Kamat S, Leon LR Jr., Eaves WB, Popma JJ, Mauri L, Donohoe D, Base CC, Rosenfield K, S.T. Investigators. Wire-interwoven nitinol stent outcome in the superficial femoral and proximal popliteal arteries: twelve-month results of the SUPERB trial. Circ Cardiovasc Interv 2015;8:e000937.
Tadros RO, Faries PL, Rocha-Singh KJ, Kim S, Malik RK, Ellozy SH, Marin ML, Vouyouka AG. The impact of sex on angioplasty and primary stenting for femoropopliteal occlusive disease: results of the DURABILITY II trial. Ann Vasc Surg. 2014;28:1–9.
Klein AJ, Chen SJ, Messenger JC, Hansgen AR, Plomondon ME, Carroll JD, Casserly IP. Quantitative assessment of the conformational change in the femoropopliteal artery with leg movement. Catheter Cardiovasc Interv. 2009;74:787–98.
Schillinger M, Minar E. Past, present and future of femoropopliteal stenting. J Endovasc Ther. 2009;16(Suppl 1):I147-152.
Early M, Lally C, Prendergast PJ, Kelly DJ. Stresses in peripheral arteries following stent placement: a finite element analysis. Comput Methods Biomech Biomed Eng. 2009;12:25–33.
Cha SH, Han MH, Choi YH, Yoon CJ, Baik SK, Kim SJ, Chang KH. Vascular responses in normal canine carotid arteries: comparison between various self-expanding stents of the same unconstrained size. Invest Radiol. 2003;38:95–101.
Kirsch EC, Khangure MS, Morling P, York TJ, McAuliffe W. Oversizing of self-expanding stents: influence on the development of neointimal hyperplasia of the carotid artery in a canine model. AJNR Am J Neuroradiol. 2002;23:121–7.
Saguner AM, Traupe T, Raber L, Hess N, Banz Y, Saguner AR, Diehm N, Hess OM. Oversizing and restenosis with self-expanding stents in iliofemoral arteries. Cardiovasc Intervent Radiol. 2012;35:906–13.
Cho H, Nango M, Sakai Y, Sohgawa E, Kageyama K, Hamamoto S, Kitayama T, Yamamoto A, Miki Y. Neointimal hyperplasia after stent placement across size-discrepant vessels in an animal study. Jpn J Radiol. 2014;32:340–6.
Schwartz RS, Huber KC, Murphy JG, Edwards WD, Camrud AR, Vlietstra RE, Holmes DR. Restenosis and the proportional neointimal response to coronary artery injury: results in a porcine model. J Am Coll Cardiol. 1992;19:267–74.
Timmins LH, Meyer CA, Moreno MR, Moore JE Jr. Mechanical modeling of stents deployed in tapered arteries. Ann Biomed Eng. 2008;36:2042–50.
Dick P, Wallner H, Sabeti S, Loewe C, Mlekusch W, Lammer J, Koppensteiner R, Minar E, Schillinger M. Balloon angioplasty versus stenting with nitinol stents in intermediate length superficial femoral artery lesions. Catheter Cardiovasc Interv. 2009;74:1090–5.
Fujihara M, Utsunomiya M, Higashimori A, Yokoi Y, Nakamura M. Outcomes of Zilver PTX stent implantation for the treatment of complex femoropopliteal artery disease. Heart Vessels. 2016;31:152–7.
Dordoni E, Meoli A, Wu W, Dubini G, Migliavacca F, Pennati G, Petrini L. Fatigue behaviour of Nitinol peripheral stents: the role of plaque shape studied with computational structural analyses. Med Eng Phys. 2014;36:842–9.
Dake MD, Ansel GM, Jaff MR, Ohki T, Saxon RR, Smouse HB, Zeller T, Roubin GS, Burket MW, Khatib Y, Snyder SA, Ragheb AO, White JK, Machan LS, Zilver PTXI. Paclitaxel-eluting stents show superiority to balloon angioplasty and bare metal stents in femoropopliteal disease: twelve-month Zilver PTX randomized study results. Circ Cardiovasc Interv. 2011;4:495–504.
Chaabane C, Otsuka F, Virmani R, Bochaton-Piallat ML. Biological responses in stented arteries. Cardiovasc Res. 2013;99:353–63.
George JC, Rosen ES, Nachtigall J, VanHise A, Kovach R. SUPERA interwoven nitinol Stent Outcomes in Above-Knee IntErventions (SAKE) study. J Vasc Interv Radiol. 2014;25:954–61.
Scheinert D, Werner M, Scheinert S, Paetzold A, Banning-Eichenseer U, Piorkowski M, Ulrich M, Bausback Y, Braunlich S, Schmidt A. Treatment of complex atherosclerotic popliteal artery disease with a new self-expanding interwoven nitinol stent: 12-month results of the Leipzig SUPERA popliteal artery stent registry. JACC Cardiovasc Interv. 2013;6:65–71.
Mohd Atan BA, Ismail AE, Taib I, Lazim Z. A review on fracture prevention of stent in femoropopliteal artery. In: IOP conference series: materials science and engineering, vol 165; 2017. p. 012006.
Litsky J, Chanda A, Stilp E, Lansky A, Mena C. Critical evaluation of stents in the peripheral arterial disease of the superficial femoral artery—focus on the paclitaxel eluting stent. Med Devices (Auckl). 2014;7:149–56.
Klein AJ, Pinto DS, Gray BH, Jaff MR, White CJ, Drachman DE, A. Peripheral Vascular Disease Committee for the Society for Cardiovascular, Interventions. SCAI expert consensus statement for femoral-popliteal arterial intervention appropriate use. Catheter Cardiovasc Interv 2014; 84:529–538.
Gokgol C, Diehm N, Nezami FR, Buchler P. Nitinol stent oversizing in patients undergoing popliteal artery revascularization: a finite element study. Ann Biomed Eng. 2015;43:2868–80.
Acknowledgements
Not applicable.
Funding
None.
Author information
Authors and Affiliations
Contributions
HL, SWR, SYC, RFY, and WGC contributed to the conceptualization, SWR, SYC, SKM and JHA contributed significantly to investigation; HL, SWR, WHK, YGR, DJO and BGC were performed methodology, HL, SYC, SKM, WWB, WYJ and WHK participated the original draft writing; HL, BGC, CUC, MJB and SHP participated the review and editing writing. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
All methods were carried out in accordance with relevant guidelines and regulations. Animal experiments were conducted at the K-BIO Osong Animal Research Center and All animal-related protocols followed the guide of ARRIVE Guidelines and approved by Animal Care and Use Review Board (No. KBIO-IACUC-2016-005) according to the guidelines by the National Institute of Health. Data collected from patients were collected by a trained study-coordinator, following a standardized case report form. The participants or their legal guardians provided written consent to participate in the study. The study protocol was approved by Medical Review Board of KUGH.
Consent to publish
Not applicable.
Competing interests
The authors have no financial conflicts of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Li, H., Rha, SW., Choi, B.G. et al. Impact of chronic outward force on arterial responses of proximal and distal of long superficial femoral artery stent. BMC Cardiovasc Disord 21, 323 (2021). https://doi.org/10.1186/s12872-021-02141-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12872-021-02141-z