Skip to main content

Table 2 Procedural characteristics in patients with a CRT-D/P (mean ± SD) (n = 21)

From: Effects of adaptive left bundle branch–optimized cardiac resynchronization therapy: a single centre experience

 

Total (n = 21)

Group 1 (n = 10)

Group 2 (n = 11)

P value

LBBA lead

    

R-wave amplitude

9.9 ± 7.2

Threshold (unipolar) (V/0.4 ms)

0.84 ± 0.17

Impedance (unipolar) (Ω)

678 ± 102

LVAT (ms)

75.2 ± 9.4

RV

DF lead

    

R-wave amplitude

23.5 ± 8.4

24.3 ± 11.8

23.0 ± 6.5

0.825

Threshold (unipolar) (V/0.4 ms)

0.82 ± 0.20

0.93 ± 0.10

0.75 ± 0.23

0.187

Impedance (unipolar) (Ω)

578 ± 147

626 ± 77

546 ± 180

0.434

CS lead

    

R-wave amplitude

18.3 ± 9.4

13.8 ± 2.6

22.1 ± 11.6

0.145

Threshold (unipolar) (V/0.4 ms)

1.0 ± 0.24

0.96 ± 0.27

1.12 ± 0.20

0.301

Impedance (unipolar) (Ω)

708 ± 134

745 ± 97

678 ± 160

0.434

CRTD (%)

19(90.5%)

9(90.0%)

10(90.9%)

1.000

Fluoroscopy Time (min)

25.2 ± 7.1

29.2 ± 8.8

21.8 ± 3.1

0.086

Procedure time (min)

135 ± 26

152 ± 31

122 ± 10

0.04

  1. LBBA left bundle branch area, LV left ventricle, RV right ventricle, DF defibrillation, CS coronary venous, LVAT left ventricular activation time