Skip to main content

Table 2 Comparison and correlation analysis of 3D-TEE and CT measurement of annulus and LVOT

From: Application of three-dimensional transesophageal echocardiography in preoperative evaluation of transcatheter aortic valve replacement

Measurement

Sample size

3D-TEE

CT

CT-TEE

P

Correlation

R

P

Systole

Annulus-Da

45

24.52 ± 2.36

25.55 ± 2.59

1.03 ± 0.98

< 0.01

0.93

< 0.01

Annulus-Dp

45

25.69 ± 2.56

26.36 ± 2.67

0.67 ± 1.13

< 0.01

0.91

< 0.01

Annulus-Mean

45

25.24 ± 2.47

25.93 ± 2.61

0.69 ± 1.20

< 0.01

0.89

< 0.01

LVOT-Dp

45

27.57 ± 3.91

28.48 ± 3.92

0.91 ± 2.19

< 0.01

0.84

< 0.01

Diastole

Annulus-Da

45

23.54 ± 2.86

24.86 ± 2.61

1.32 ± 1.46

< 0.01

0.86

< 0.01

Annulus-Dp

45

25.11 ± 2.85

26.02 ± 2.70

0.91 ± 1.28

< 0.01

0.90

< 0.01

Annulus-Mean

45

24.14 ± 2.90

25.44 ± 2.63

1.30 ± 1.63

< 0.01

0.83

< 0.01

LVOT-Dp

45

27.37 ± 4.73

29.79 ± 4.50

2.42 ± 3.59

< 0.01

0.70

< 0.01