Table 4 Area Under the Curve (AUC) for the Receiver Operating Curves (ROC) comparing the MissionHb® and Urit 12® meters to the reference analyser

From: Evaluation of the accuracy of two point-of-care haemoglobin meters used in sub-Saharan African population: a cross-sectional study

Meters

Area under curve (AUC) [95% CI]

Reference Analyser

0.9776 [09624–0.9929]

MissionHb®

0.9161 [0.8764–0.95586]

Urit 12®

0.9009 [0.8601–0.9416]