Skip to main content

Table 3 Procedural characteristics

From: Novel sirolimus-eluting stent Prolim® with a biodegradable polymer in the all-comers population: one year clinical results with quantitative coronary angiography and optical coherence tomography analysis

Parameter

n = 204 (%)

Device success

202 (99.0)

No of stents per patient

1.17

Predilatation

122 (59.8)

Postdilatation

56 (27.5)

Nominal stent diameter [mm]

3.22 ± 0.52

Nominal stent length [mm]

18.41 ± 5.41

Stent maximal inflation pressure [atm]

15.23 ± 2.54

Balloon to artery ratio

1.10 ± 0.07

Additional stent implantation due to dissection

14 (6.8)

Additional stent implantation due to lesion length

16 (7.8)

Vascular access radial/femoral

189 (92.6)/15 (7.4)

Guiding catheter 6 F/7 F

202 (99.0)/2 (1.0)