Table 3 Detailed diagnostic performance of the wearable dynamic ECG recorder for AF screening in a supine position

From: Diagnostic performance of a wearing dynamic ECG recorder for atrial fibrillation screening: the HUAMI heart study

 

Unable to unclassified as TPTN

Unable to unclassified as FPFN

Accuracy, %(95% CI)

96.49% (91.03–98.92%)

94.74% (88.76–97.80%)

Sensitivity, %(95% CI)

92.45% (81.64–97.52%)

88.68% (77.06–95.07%)

Specificity, %(95% CI)

100% (92.91–100%)

100% (92.91–100%)

PPV, %(95% CI)

100% (91.32–100%)

100% (90.98–100%)

NPV, %(95% CI)

93.85% (84.78–98.02%)

91.04% (81.48–96.16%)

  1. ECG electrocardiogram, AF atrial fibrillation, TP true positives, TN true negatives, FP false positives, FN false negative, PPV positive predictive value, NPV negative predictive value