Table 2 Indicators of a 24-h BP monitoring in the brachial artery and aorta in women at various climacteric phases

From: Evaluating daily profile of central aortic pressure and reflected pulse wave parameters in climacteric women

Indicators

Group 1 (n = 168)

Group 2 (n = 216)

p1-2

SBP—24-h period, mmHg

115.7 ± 11.3

126.2 ± 13.3

 < 0.001

SBP—day, mmHg

118.5 ± 11.8

130.7 ± 13.9

 < 0.001

SBP—night, mmHg

106.5 ± 12.1

116.8 ± 14.9

0.001

PBP—24-h period, mmHg

43.7 ± 6.2

49.3 ± 9.5

0.014

PBP—day, mmHg

45.6 ± 6.5

52.0 ± 10.0

0.032

PBP—night, mmHg

40.9 ± 6.2

47.4 ± 9.6

0.013

CASP—24-h period, mmHg

108.4 ± 10.8

118.2 ± 12.0

 < 0.001

CASP—day, mmHg

109.7 ± 10.7

119.6 ± 12.3

0.001

CASP—night, mmHg

99.3 ± 10.1

110.5 ± 15.6

0.012

CAPP—24-h period, mmHg

34.1 ± 5.8

39.9 ± 8.0

 < 0.001

CAPP—day, mmHg

34.4 ± 6.0

40.6 ± 8.4

0.002

CAPP—night, mmHg

32.3 ± 5.4

39.4 ± 11.9

0.001

PPA—24-h period, mmHg

9.6 ± 2.1

9.5 ± 2.2

0.326

PPA—day, mmHg

11.2 ± 2.2

11.4 ± 2.2

0.188

PPA—night, mmHg

8.6 ± 1.9**

8.0 ± 1.7**

0.002

AIxao—24-h period, %

22 (15–28)

28 (20–35)

0.004

AIxao—day, %

20 (12, 25)

27.5 (22, 35)

0.018

AIxao—night, %

23.5 (17, 28)*

32 (22, 38)*

0.001

  1. The data are presented in the form of M ± SD or Me (LQ, UQ); p1-2 represents the significance of differences between the women of Groups 1 and 2; asterisks show the significance level of the differences between the daytime and nighttime monitoring periods (*p < 0.05; **p < 0.01); SBP stands for systolic blood pressure; PBP indicates pulse blood pressure; CASP is central aortic systolic pressure; CAPP is central aortic pulse pressure; AIxao is aortic augmentation index; PPA represents pulse pressure amplification