Skip to main content

Table 3 Comparison of the pacing parameters between groups A and B

From: Effect of implantation site of the His bundle pacing leads on pacing parameters: a single-center experience

Time schedule Capture (V/0.5 ms) Sensing value (mV) Impedance
A (n = 13) B (n = 16) A (n = 13) B (n = 16) A (n = 13) B (n = 16)
During operation 1.20 ± 0.34 1.14 ± 0.43 1.87 ± 0.82 4.53 ± 1.37* 626 ± 91 690 ± 123
1 month 0.69 ± 4.44** 0.81 ± 0.39* 1.95 ± 0.76 4.69 ± 1.38* 529 ± 163 528 ± 117
1 year 0.92 ± 0.80 0.98 ± 0.59 1.88 ± 0.75 4.59 ± 1.42* 522 ± 181 574 ± 178
  1. *p < 0.05 versus group A
  2. **p < 0.05 1 month versus during operation