Table 4 Details on the ICD programming divergences from the recommended settings (presented as Figures 5 and 6 in Appendix)

From: Arrhythmic episodes in patients implanted with a cardioverter-defibrillator – results from the Prospective Study on Predictive Quality with Preferencing PainFree ATP therapies (4P)

 

TOTAL

(n = 185)

SC-ICD

(n = 44)

DC-ICD

(n = 48)

CRT-D

(n = 93)

Primary prevention (pts)

121

19

25

77

 FVT detection- ON [n, (%)]

19 (10.2)

1 (2.3)

4 (8.3)

14 (15.1%)

 VT detection- ON [n, (%)]

90 (74.4)

15 (78.9)

18 (72.0)

57 (64.0)

 PR Logic/ Wavelet- OFF or Monitor [n, (%)]

18 (14.9)

3 (15.8)

2 (8.0)

13 (16.9)

 Onset ON or OFF [n, (%)]

49 (40.5)

12 (53.1)

9 (36.0)

28 (36.4)

 T-Wave- OFF [n, (%)]

6 (5)

3 (15.8)

0

3 (3.9)

 RV-Lead Noise [n, (%)]

3 (2.5)

2 (10.5)

0

1 (1.3)

 ATP before charging

8 (6.6)

1 (5.3)

2 (8.0)

5 (6.5)

 FVT- ATP & shocks

121 (100)

19 (100)

25 (100)

77 (100)

 VT- ATP & shocks

121 (100)

19 (100)

25 (100)

77 (100)

Secondary prevention (pts)

64

25

23

16

 FVT detection- ON [n, (%)]

19 (29.7)

8 (32.0)

9 (39.1)

2 (12.6)

 VT detection- OFF [n, (%)]

11 (17.2)

6 (24.0)

5 (21.7)

0

 PR Logic/ Wavelet- OFF or Monitor [n, (%)]

4 (6.3)

3 (12.0)

0

1 (6.3)

 Onset ON or OFF [n, (%)]

32 (50.0)

10 (40.0)

14 (60.8)

8 (50.0)

 ATP before charging

8 (12.5)

3 (12.0)

2 (8.7)

3 (18.8)

 FVT- ATP & shocks

121 (100)

19 (100)

25 (100)

77 (100)