Skip to main content

Table 3 Procedural data

From: The safety of introducing a new generation TAVR device: one departments experience from introducing a second generation repositionable TAVR

 

Lotus (n = 54)

Sapien (n = 47)

p-value

Procedural time (min)

82.9 (31.0)

118.7 (99.7)

0.0169

Fluoroscopy time (min)

29.7 (12.3)

26.2 (13.3)

0.1637

Contrast (mL)

92.6 (31.0)

99.3 (32.1)

0.2856

General anesthesia

39.0% (21)

100% (47)

0.0000

Pre-dilatation

59.3% (32)

83% (39)

0.0089

Post-dilatation

0% (0)

27.7% (13)

0.0000

Rapid Pacing

22.2% (12)

100% (47)

0.0000

Per-op bleeding (mL)

105.1 (159.9)

202.8 (287.4)

0.0355

Heart-lung machine (unplanned)

1.9% (1)

8.5% (4)

0.1927

New pacemaker

15.1% (8)

6.8% (3)

0.2172

Aortic valve malpositioning

0% (0)

0% (0)

 

Valve migration

0% (0)

0% (0)

 

Valve embolization

0% (0)

0% (0)

 

Ectopic valve deployment

0% (0)

0% (0)

 

TAV-in-TAV deployment

0% (0)

0% (0)

 
  1. Procedural data in all 54 Lotus procedures, but new pacemaker reported for the 53 patients