Table 2 The PWV median errors for different number of time frames/temporal resolution based on 210 computer phantoms

From: Required temporal resolution for accurate thoracic aortic pulse wave velocity measurements by phase-contrast magnetic resonance imaging and comparison with clinical standard applanation tonometry

Time frames

Temporal resolution [ms]

Median error [m/s]

Median error [%]

20

50.0

2.36

21.3

21

47.6

1.51

17.2

22

45.5

0.72

10.4

23

43.5

0.16

9.1

24

41.7

-0.02

6.5

25

40.0

-0.39

4.4

26

38.5

-1.61

10.8

27

37.0

-2.74

18.1

28

35.7

-2.39

21.3

29

34.5

-1.83

16.2

30

33.3

-1.29

11.4

31

32.3

-0.76

6.7

32

31.3

-0.42

5.1

33

30.3

-0.20

2.6

34

29.4

-0.11

1.3

35

28.6

0.10

0.9

36

27.8

0.48

3.6

37

27.0

0.46

4.1

38

26.3

0.30

3.0

39

25.6

0.23

2.4

40

25.0

0.17

1.8

  1. Errors are expressed as median errors for all velocities in the range of 2–20 m/s