Skip to main content

Table 3 Standard deviation of the mean differences between the two measurement periods

From: Reproducibility of wrist home blood pressure measurement with position sensor and automatic data storage

 

systolic

diastolic

 

SD

95% CI

SD

95% CI

Home BP

3.81

3.34 – 4.44

2.77

2.43 – 3.22

24-h ABPM

7.83

6.90 – 9.17

4.09

3.59 – 4.77

Office BP

8.10

7.10 – 9.43

4.76

4.18 – 5.54

p- value

    

Home vers. 24-h ABPM

p < 0.001

p = 0.127

Home vers. Office BP

p < 0.001

p < 0.001

Office vers 24-h ABPM

p = 0.80

p = 0.10

  1. SD: Standard deviation of mean differences. 95% CI: 95% confidence interval.