Table 1 Mean systolic and diastolic BPs

From: The BpTRU automatic blood pressure monitor compared to 24 hour ambulatory blood pressure monitoring in the assessment of blood pressure in patients with hypertension

Average of the blood pressures measured at the last three office visits 150.8 (SD10.26) / 82.9 (SD 8.44)
BpTRU initial reading 150.0 (SD21.33) / 83.3 (SD12.00)
BpTRU Average 140.0 (SD17.71) / 79.8 (SD 10.46)
24 hour daytime average 141.5 (SD 13.25) / 79.7 (SD 7.79)