Skip to main content

Table 2 Intra-observer concordances of ABI measurements

From: Validity, intra- and inter-observer reliability of automated devices for the assessment of ankle brachial index using photo-plethysmography

Method

CCC

95%-CI of CCC

Bias

p-value

Correlation

VI

0.65

0.46

0.78

0.015

0.29

0.65

VI_O

0.68

0.51

0.80

0.001

0.95

0.68

VI_E

0.65

0.47

0.78

0.006

0.65

0.66

VE

0.65

0.50

0.76

-0.011

0.43

0.66

VE_O

0.75

0.59

0.85

-0.004

0.75

0.75

VE_E

0.76

0.57

0.87

-0.002

0.89

0.76

VE_D

0.57

0.27

0.78

0.021

0.12

0.59

  1. Agreement of first and second ABI measurement for automated methods. We present concordance correlation coefficients, corresponding confidence intervals, bias between first and second measurements (values less than zero mean lower values at first measurement), p-value of bias and correlation coefficient of first and second measurement.
  2. VI/VE = Vicorder/Vascular Explorer standard measurements, VI_O/VE_O = Vicorder/Vascular Explorer corrected by observer, VI_E/VE_E = Vicorder/Vascular Explorer corrected by examiner, VE_D = deflation method of Vascular Explorer.