Chronic lymphocytic leukemia skin infiltration mimicking an ICD pocket infection: a case report

Abstract

Background

We are presenting a case report on an unreported and unusual cutaneous manifestation of chronic lymphocytic leukemia in a patient with an implantable cardioverter-defibrillator (ICD).

Case presentation

A 65-year-old man with a history of chronic lymphocytic leukemia (CLL), previously treated with chlorambucil, was referred in October 2013 for extraction of a single chamber ICD due to a suspected device-related infection in the pulse generator area (left-hand side of Fig. 1). The ICD system (Current VR, St. Jude Medical, USA) had been implanted in November 2009. The patient complained of painless erythema with pruritus in the pocket area. Inflammatory blood parameters were C-reactive protein 17.3 mg/L and leucocytes 29.0 × 109/L. Due to the atypical appearance of the pocket area we did not extract the device. Instead, we created an exploratory excision in the skin induration, which had been present for approximately 6 weeks, and conducted a microbiological and histological examination. All cultivation examinations were negative. However, we did histologically show skin infiltration by CD-5 positive low-grade B-cell chronic lymphocytic leukemia/small lymphocytic lymphoma (B-CLL/SLL). Re-initiation of chemotherapy was not necessary and the skin induration completely disappeared within 2 months (right-hand side of Fig. 1).

Conclusions

Complete removal of an ICD system carries considerable risk. In patients with a history of hematological disease, it is crucial to exclude cutaneous manifestations of the disease prior to device removal.

Peer Review reports

Background

We are presenting a case report on an unreported and unusual cutaneous manifestation of a chronic lymphocytic leukemia in a patient with an implantable cardioverter-defibrillator.

Case presentation

A 65-year-old man with a history of chronic lymphocytic leukemia (CLL), previously treated with chlorambucil, but currently not on active treatment, was referred in October 2013 for extraction of a single chamber implantable cardioverter-defibrillator (ICD) because of a suspected device-related infection in the pulse generator area (left-hand side of Fig. 1). The ICD system (Current VR, St. Jude Medical, USA) had been implanted in November 2009 for primary prevention of sudden cardiac death in the setting of idiopathic dilated cardiomyopathy with low left ventricular ejection fraction (25%). The patient complained of painless erythema with intense pruritus in the pocket area. Inflammatory blood parameters were C-reactive protein 17.3 mg/L (ULN < 5), procalcitonin 0.2 ng/ml (ULN < 0.5) and leucocytes 29.0 × 109/L (ULN < 10), and with 84% lymphocytes. The patient had no fever and his body temperature was in range of 36.5–36.9 °C. Due to the atypical appearance of the pocket area we deferred extracting the device. Instead, we created an exploratory excision of the skin induration, which had been present for approximately 6 weeks, and conducted a microbiological and histological examination. Both aerobic and anaerobic cultivation examinations were negative. However, skin infiltration by CD-5 positive low-grade B-cells was found, which suggested a diagnosis of chronic lymphocytic leukemia/small lymphocytic lymphoma (B-CLL/SLL). Re-initiation of chemotherapy was not necessary because of the disease was considered stable and the skin induration spontaneously and completely disappeared (right-hand side of Fig. 1) over following 2 months and no relapse of symptoms has been observed till present time.

Fig. 1
figure 1

Left-hand side: Painless erythema and mild induration over the device pocket suspicious of device-related infection. Right-hand side: skin induration spontaneously disappeared over following two months

Discussion

Infectious complications, following cardiac implantable electronic device (CIED) implantation have steadily increased since it first entered clinical practice and are associated with significant mortality. Important host-related risk factors presented in a recently published meta-analysis included: diabetes mellitus, renal disease, chronic obstructive pulmonary disease, corticosteroid use, malignancy, heart failure, and anticoagulant drug use. Procedure-related risk factors included lack of antibiotic prophylaxis, replacement/revision procedures, non-infectious post-operative complications (including dislodged leads and hematoma), temporary pacing, and procedure duration. Average device infection rates have been reported between 1–1.3% [1].

Pacemaker component allergy is a relatively uncommon cause of erythema and pain at the pacemaker implantation site. Diagnosis of a pacemaker component allergy first requires awareness of the problem and then thorough allergy testing with the appropriate allergy test kit provided by the manufacturer. Pacemaker component allergy can present as painful erythema with pruritus covering a large area around the device and may occur months after the implantation procedure. Therapy with a topical corticoid may permanently resolve the skin reaction; otherwise, a more extensively coated or a gold-plated device may need to be implanted in place of the original device [2].

A very rare condition, which can mimic a pacemaker pocket infection, is breast carcinoma. It has been reported in both male and female patients. A firm lesion with signs of skin erosion can be observed near the pacemaker pocket; however, C-reactive protein, white-cell count, and other routine blood tests can be normal [3].

Conclusions

Removal of CIEDs and all associated leads carries considerable risks. Therefore, in patients with a history of hematological disease, it is crucial to exclude cutaneous manifestations of the disease prior to device removal. Other conditions mimicking pocket infection include breast cancer or pacemaker component allergy.

Abbreviations

B-CLL:

B-cell chronic lymphocytic leukemia

CIED:

Cardiac implantable electronic device

CLL:

Chronic lymphocytic leukemia

ICD:

Implantable cardioverter-defibrillator

SLL:

Small lymphocytic lymphoma

ULN:

Upper limit of normal

References

  1. Polyzos KA, Konstantelias AA, Falagas ME. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace. 2015;17(5):767–77.

    Article  PubMed  Google Scholar 

  2. Citerne O, Gomes S, Scanu P, Milliez P. Painful Eczema mimicking pocket infection in a patient with an ICD: a rare cause of skin allergy to nickel/cobalt alloy. Circulation. 2011;123(11):1241–2.

    Article  PubMed  Google Scholar 

  3. De Mattia L, Brieda M, Dametto E. A carcinoma of the breast mimicking a pacemaker pocket infection. Europace. 2011;13(2):220.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Pavla VITKOVA, Department of Pathology, Ceske Budejovice Hospital, B. Nemcove 585/54, 370 01 Ceske Budejovice, Czech Republic. PV performed the histological examination.

Cardiocenter Ceske Budejovice Foundation for financial support of manuscript publication.

Funding

Foundation Cardiocenter Ceske Budejovice financially supported the publication costs of the manuscript.

Availability of data and materials

Not applicable.

Authors’ contributions

MS: patient followed-up and contributed to writing the manuscript. AB: case senior cardiologists, performed the procedure and contributed to writing the manuscript, performs all duties of the corresponding author. IV: interpreted patient findings regarding the hematological disease. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Written informed consent for the publication of all the information contained in the case report and any accompanying images has been obtained from the patient.

Ethics approval and consent to participate

Not applicable.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to A. Bulava.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Snorek, M., Bulava, A. & Vonke, I. Chronic lymphocytic leukemia skin infiltration mimicking an ICD pocket infection: a case report. BMC Cardiovasc Disord 17, 89 (2017). https://doi.org/10.1186/s12872-017-0522-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12872-017-0522-5

Keywords